User's Manual

Table Of Contents
Technical Data
14
Default Programs
Parameter/Function
Factory- Standard- Safe
Settings Program Program
Home Monitoring
Off Off Off
Mode
DDD DDDR VVI
Autom. m. conversion
Off Off
Basic rate
60 ppm 60 ppm 70 ppm
Hysteresis
Off Off Off
Repetitive hysteres.
——Off
Scan hysteresis
——Off
Night program
Off Off Off
Upper rate (UTR)
160 ppm 160 ppm
Dynamic AV delay
medium medium
Ventric. blanking time
24 ms 24 ms
Magnet effect
Auto Auto Auto
Pulse amplitude A
V
3,6 V 3,6 V
3,6 V 3,6 V 4,8 V
Pulse width A
V
0,4 ms 0,4 ms
0,4 ms 0,4 ms 1,0 ms
Sensitivity A
V
1,5 mV 1,5 mV
2,5 mV 2,5 mV 2,5 mV
Refractory time A
V
425 ms 425 ms
250 ms 250 ms 300 ms
ARP
0 ms 0 ms