User's Manual

Table Of Contents
5
Transmission of Message
Transmission of Message
The Home Monitoring function can be switched ON or
OFF with the programming device. If the Home Moni-
toring function is active, the transmission of the
implant data can be triggered as follows:
Periodic message at predefined time intervals
Message activated by the patient
The attending physician himself decides if the patient
should trigger a transmission by programming this
paramater ON or OFF. The patient activated message
does not effect the programmed periodic message.
Note: Please note that with the BA03 DDDR, the Home Mon-
itoring function is only possible in following pacing
modes: DDD, DDDR, DDI, DDIR.
Periodic Message
The time and interval (Monitoring Interval) of the peri-
odic message are programmable. For the periodic
message, a time between 0:00 and 24:00 can be set.
Generally, it is recommended to select a time between
0:00 and 4:00. For every Monitoring Interval, a data
string is generated in the implant and the transmission
is activated.
Patient Activated Message
Application of a magnet over the implant activates a
transmission. The attending physician should inform
the patient in detail about the handling and the physical
symptoms that suggest a magnet application by the
patient.
Caution:
The magnet effect must be programmed "synchro-
nous" if the attending physician enables the patient to
transmit messages.