User's Manual

Table Of Contents
Functions of Home Monitoring
8
Functions of Home Monitoring
Programmable Parameters
The functions of Home Monitoring and their
parameters can be set with the programmers EPR
1000
plus
and TMS 1000
plus
. Therefore, the correspond-
ing software SWM 1000/B-KT0.0 is necessary
The functions of Home Monitoring are available only in
following pacing modes:
DDD, DDDR, DDI, DDIR.
When using the Home Monitoring system, the trans-
mission interval must be selected. This monitoring
interval can be from one to 30 days. If the standard set-
ting is selected, data will be transmitted daily. For the
periodic message, a time between 0:00 and 24:00 is
programmed. It is recommended that a time between
0:00 and 4:00 is selected, as preset in the standard pro-
gram.
The patient activated message can also be pro-
grammed ON or OFF. This option is deactivated in the
standard program