Cardiac Airbag Technical Manual 39
2.6.5 System Status
Various device parameters can be monitored through the Status
section of the programmer screen. (See Figure
4) Displayed
data includes ICD information, charge circuit parameters,
capacitor reformation data, battery status, and lead information.
The system status screen presents a large variety of information
about the Cardiac Airba
g ICDs including:
• Serial number (always displayed after interrogation)
(End of Service)
• Last charge event
- Date
- Energy
- Charge time
• Total number of charges
• Last R-wave measurements
• Last pacing lead impedance (ventricle)
• Last pacing threshold measurement with pulse width
(ventricle)
• Last shock impedance measurement and date
• Software release
• Device status
• Battery status
- BOL (Begin of Life)
- ERI (Elective Replacement Indication)
- EOS