Cardiac Airbag Technical Manual 59
Figure 11. Manual Cap Reform
Acc for lead and device-basedessing the TMS 1000
PLUS
ing
1. Press Options button.
2. In the Programmer area, press Implant List button
to return to the main programmer screen.
3. Select TMS 1000 from the
test
Implant List.
Connecting the ventricular lead for testing
1. Connect the terminal pins of the ventricular lead to
the appropriate port of the PK-44 Cable.
Pace/Sense pin = PA-2 block
Shock pins = PA-3 block