64 Cardiac Airbag Technical Manual
DFT testing
1. Select method to induce ventricular fibrillation with
the ICD: Shock on T wave or HF burst.
Press Print button to be
2. gin printing induction.
3. Press Start VT/VF Induction (See Figure
14) and
OK to induce VF.
4. Observe the IEGM, markers, and ICD charging
indicator on the programmer screen.
Figure 14. VF Induction Test
5. After successful delivery of therapy, press Print to
stop printing. (See Figure
15)