Technical Data
15
4 Appendix
Technical Data
General characteristics for the device and the power adapter (configured as an medical electrical system)
Category Design
Operating mode Continuous operation
Temperature range for operation +10°C ...+33°C/50°F...91°F
Temperature range for storage 0°C ...+50°C/32°F...122°F
Relative humidity 20%...75%, no condensation
Atmospheric pressure 700...1060hPa
Operation at altitudes Up to 3,000 m
Power supply to the device Operation with power adapter (FW8000M/12)
Physical properties
Category Design
Dimensions (WxDxH) 120x150x42.8mm
Weight with programming head 230g
Longevity
Category Design
Longevity 2years
Programming head (BIOwand PGH)
Category Design
Applied part classification BF
Dimensions (WxDxH) 102x132x37mm
PGH cable 2.8m
Frequency band 9–315kHz
Operating frequency 32–64kHz
Maximum transmitter field strength < -36dBµA/m
Max.peak@10m
Modulation OOK
Data rate Up to 64kbit/s