Technical Manual

Safety during Use
Operating Conditions
5
WARNING
Risk of electromagnetic interference through the use of portable RF communication equipment
If portable RF communication devices (including peripheral devices such as antenna cables and
external antennae) are operated closer than 30cm (12inches) from this device, this can result in a
reduction in its performance. This applies even when using associated cables.
When operating portable RFcommunication devices (including peripheral devices such as
antenna cables and external antennae), keep such devices at a distance of at least 30cm
(12inches) from this device.
Only operate the device in rooms that fulfill the following conditions:
No danger of explosion
Suitable for medical purposes
ClassI power outlet
Place the device and its components on a flat, dry surface. It should be placed so that it cannot slip,
even while the cables are connected. Also ensure that it is protected from liquids and moisture, and
that the patient only comes into contact with the programming head. In addition to this, ensure that the
power plug of the power adapter is easily accessible and can be pulled out of the outlet at all times. Do
not touch any plug connections such as Redel connectors and the patient at the same time.
Power supply
The device is operated via the provided power adapter. The electrical outlet must fulfill the following
conditions:
The mains installation at least fulfills the requirements of IEC60364‑7‑710:2002group1.
The power plug of the power adapter goes directly into a permanently installed mains supply
outlet. No portable multiple socket outlets or extension cables should be used.
The power plug of the power adapter must be easily accessible at all times in order to be able to
immediately disconnect the device from the mains supply.
When used in combination with other devices, no portable multiple socket outlets may be used.
Only power supply adapter authorized by BIOTRONIK may be used.
To disconnect the device from the mains supply, pull its power plug out from the outlet.
Cable and plug connections
Replace the device or its components immediately if there is any visible, even minor damage to the
cables.
Lay all cables in such a way that they pose no danger of tripping over them and that any tensile
forces that may occur can be safely buffered.
Ensure that the contacts of all connections and plugs are clean. Dirty contacts can lead to signal
distortions.
Ensure that there is no condensation on the plugs or in the connector ports. If condensation is
present, dry it before use.
Do not force plugs into the connector ports and do not pull on the cable to release the lock when
disconnecting the plugs.