14 Switching Off the System
Switching Off the System
Normal shut-down When you press the On/Off button quickly, the
Operation Module switches to standby mode within a
few seconds.
The data are saved.
The screen goes black.
The battery level of the rechargeable battery is
checked; if it is fully charged, the Operation
Module and Docking Station will switch off
automatically after 30 seconds.
If the Operation Module is connected, the system
connected to the power supply and the battery is no
longer fully charged, it will be automatically recharged
(see also „Automatic Battery Recharge“ on page 15).
The screen goes black and the fan continues to
run.
The system shuts down completely only after the
battery has been recharged.
Note: After a normal shut-down, the ICS is quickly ready for
operation again: Depending on the pre-defined
setting, a restart takes 30 or 60 seconds.
Forced shut-down If the software does not respond to brief pressure on
the button for normal shut-down, you can force the
system to shut down.
When you press the On/Off button for 3 seconds, the
Operation Module switches itself off immediately.
Any unsaved data will be lost.
The battery will not be automatically recharged.
The subsequent restart takes about 60 seconds.
WARNING! To switch off the device in an emergency, unplug the
device. In an emergency, the device cannot be
switched off effectively using the On/Off button on the
Operation Module, because there is a delay in
switching off. The ON switch on the Docking Station is
used only for switching on the device.