User Manual

37 Technical Data
Miniclinic
Stimulation modes Single- and multi-chamber
Pacing rate [ppm] 30 … 180; ±2
Period
[ms]
333 … 2000; 2 ±4
Pulse width (A+V)
[ms]
0,1 … 2,5; 0,05 ±0,05
AV conduction time
[ms]
50 … 300; 2 ±4
Miniclinic
Trigger level [mV]
2 … 150; 2 ±4
PGH Programming Head
ICS 3000PGH Spiral cable, extendable to approx. 2.30 meters
ICS 3000PGH Straight cable, 2.1 meters
Straight cable, 2.9 meters
Dimensions [mm] 142*97*42 (W*H*D)
Weight [kg] 0,5
PGH connection Redel plug, 14-pin
Protection degree IP 30
Protection degree
for application part
Type B
Magnetic flux
density [mT]
>2.0 at 60 mm distance
>20.0 at 10 mm distance
30.0 at 0 mm distance
Docking Station: General Information
Dimensions [mm] 284*103*322 (W*H*D)
Weight [kg] 3,8
Cooling with fan
Fan control temperature-controlled