User Manual

45 Electromagnetic Compatibility
Note: At 80 MHz and at 800 MHz, the higher frequency range
applies.
Magnetic field at the
supply frequencies
(50/60 Hz)
According to
IEC 61000-4-8
3 A/m Same as test
level
Ensure that the magnetic
field strengths are at
levels characteristic of a
location in a typical
commercial and/or
hospital environment.
Test of Interference
Resistance
Test Level
According to
IEC 60601-1-2
Compliance Guidelines for the
Electromagnetic
Environment
Test of Interference
Resistance
Test Level
According to
IEC 60601-1-2
Compliance Guidelines for the
Electromagnetic
Environment
Conducted
RF interference
According to
IEC 61000-4-6
3V
eff
3 V Maintain safety distance
of mobile radio equipment
to the ICS 3000; see
page 5
The field strength of
stationary transmitting
devices must be
measured on site and
must be lower than the
compliance level at all
frequencies: consider any
studies done on site.
The field strength must be
lower than 3 V/m over the
frequency range of
150 kHz to 80 MHz.
Radiated
RF interference
According to
IEC 61000-4-3
3V/m
80 MHz to
2.5 GHz
3V/m