Product Description
Diagnostic and Therapy Functions
Resynchronisation therapy
Triple-chamber devices have functions to configure different VV delays in order to
resynchronize the ventricles.
Capture Control is also available for the left ventricle with automated tracking of
the pacing threshold or automatic threshold monitoring (ATM) for trend
analysis.
To ensure that no additional surgery is necessary in case of a left-sided increase
of pacing threshold or undesired phrenic nerve stimulation, different pacing
polarities can be set for the left ventricular lead with a triple-chamber device.
Up to 13
vectors can be used with the HF QP device type.
With the QP device type, the LV vector test provides a fast measurement of the
pacing threshold, the phrenic nerve pacing threshold and the pacing
impedance. The relative influence on the service time is also displayed. The
measurement results are evaluated automatically so that the optimal pacing
polarity can be set.
The short RV-LV conduction test supports also the selection.
An additional diagnostic function with biventricular pacing: Variability of the
heart rate, patient activity, and thoracic impedance are monitored on a continual
basis.
Programs
There are two types of therapy programs:
Default parameters are offered for the most common indications
(ProgramConsult function).
Individual settings can be saved in 3 individual therapy programs.
ProMRI devices recognize
magnetic resonance imaging
devices
The static magnetic field of magnetic resonance imaging devices is reliably
recognized with the aid of a sensor. This sensor can be activated for a maximum of
14 days using the MRI AutoDetect function during an interrogation.
If the patient comes near a magnetic resonance imaging device within the time set,
the implanted device recognizes the static magnetic field and automatically
activates the preset MRI program. Reprogramming to the permanent program
occurs also automatically after leaving the imaging device.
Home Monitoring functions
The device automatically sends information to the transmitter once a day. In
addition to this, test messages can be initiated using the programmer. Important
medical information includes, among others, the following:
Ongoing atrial and ventricular arrhythmia
Parameters relevant to leads in the atrium and ventricle: thresholds, sensing
amplitudes, impedances
Current statistics on bradycardia therapy
Individually adjustable timing interval for device messages which provide
additional information pertaining to the device messages
IEGM online HD
®
with up to 3 high definition channels
Transmission of these IEGM recordings with device messages