User Manual

Table Of Contents
15
Implantation
Implantation Procedure
3 Implantation
Implantation3417803Technical ManualEdora 8
Implantation Procedure
Having parts ready
The following parts that correspond to the requirements of the EC Directive
90/385/EEC are required:
Device with screwdriver from BIOTRONIK
BIOTRONIK leads and lead introducer set
Single-chamber device: unipolar or bipolar lead for the right ventricle
Dual-chamber device: one unipolar or bipolar lead each for the atrium and
for the right ventricle
Triple-chamber device: an additional unipolar, bipolar, or quadripolar LV
lead
Approved connections are IS-1 and IS4: Use only adapters approved by
BIOTRONIK for leads with diff
erent connections or leads from other
manufacturers.
BIOTRONIK programmer (with integrated wandless telemetry or with separate
SafeSync Module) and approved cables
External multi-channel ECG device
Keep spare parts for all sterile components.
Keeping an external
defibrillator ready
In order to be able to respond to unforeseeable emergencies or possible technical
failures of the device:
Keep an external defibrillator and paddles or adhesive electrodes ready.
!
!
WARNING
Inadequate therapy due to defective device
If an unpacked device is dropped on a hard surface during handling, electronic
parts could be damaged.
Unpacking the device
Peel the sealing paper off of the outer blister at the marked position in the
direction indicated by the arrow. The inner blister must not come into contact
with persons who have not sterilized their hands or gloves, nor with non-sterile
instruments.
Use the gripping tab on the inner blister to remove it from the outer blister.
Peel the sealing paper off of the sterile inner blister at the marked position in
the direction indicated by the arrow.
Note: The device is disabled on delivery and can be implanted immediately after
unpacking without manual activation.
Checking parts
Damage to any of the parts can result in complications or technical failures.
Check for damage before and after unpacking all parts.
Replace damaged parts.
Use a replacement device.
Return the damaged device to BIOTRONIK.