User Manual
Table Of Contents
- Edora 8
- Table of Contents
- 1 Product Description
- 2 General Safety Instructions
- 3 Implantation
- Implantation Procedure
- Having parts ready
- Keeping an external defibrillator ready
- Unpacking the device
- Checking parts
- Implantation site
- Overview: Implanting
- Avoid damage to the header
- Preventing short circuits in the header
- Keeping distance between leads
- Connecting the lead connector to the device
- Applying the programming head
- Establishing wandless telemetry
- Auto-initialization
- Behavior during auto- initialization
- Precautionary Measures while Programming
- Checking the device system
- Performing standard tests and monitoring the patient
- Do not interrupt wandless telemetry during a treatment
- Cancelling telemetry
- Avoiding critical parameter settings
- Manually setting lead polarity
- Setting sensing
- Setting the sensitivity
- Preventing device-induced complications
- Preventing conduction of atrial tachycardia
- Phrenic nerve stimulation that cannot be terminated
- Avoiding risks in the case of exclusive left ventricular pacing
- If an ICD is implanted at the same time, do not permit unipolar pacing
- Recognizing lead failure
- Consider power consumption and service time
- Magnet Response
- Follow-up
- Patient Information
- Replacement Indications
- Explantation and Device Replacement
- Implantation Procedure
- 4 Parameters
- 5 Technical Data
27
Parameters
Timing
4 Parameters
Parameters4417803Technical ManualEdora 8
Note: Unless described separately, information for device type HF also applies to
device type HF QP.
Timing
Basic rate day/night
Rate hystereses
AV delay
AV hystereses
Parameter Range of values Standard SR DR HF
Basic rate 30 ... (5) ... 100 ... (10)
... 200 bpm
60 bpm x x
50 bpm x
Night rate OFF; 30 ... (5) ... 100 ... (10)
... 200 bpm
OFF x x x
Night begins 00:00 ... (10 min) ...
23:50 hh:mm
—xxx
Night ends
Parameter Range of values Standard SR DR HF
Hysteresis OFF; -5 ... (-5) ... -25 ... (-20)
... -65 bpm
OFF x x x
Repetitive/ search
cycles
OFF; ON OFF x x x
Parameter Range of values Standard SR DR HF
AV delay Low; Medium; High; Fixed;
Individual
Low x x
20 ... (5) ... 350 ms
(in 6 rate ranges)
180-170-
160-150-
140 ms
x
CLS and all HF modes:
20 ... (5) ... 350 ms
(in 6 rate ranges)
150-140-
130-120-
120 ms
xx
Sense compensation OFF; -10 ... (-5) ... -120 ms -45 ms x x
Parameter Range of values Standard SR DR HF
AV hysteresis mode OFF; Positive; Negative
HF when setting RV: IRSplus
OFF x x
Positive modes:
AV hysteresis
70; 110; 150; 200 ms 70 ms
CLS modes:
110 ms
xx
Negative modes:
AV hysteresis
10 ... (10) ... 150 ms 50 ms x x
AV repetetive /
scan cyles
OFF; ON ON x x