User Manual
Table Of Contents
- Edora 8
- Table of Contents
- 1 Product Description
- 2 General Safety Instructions
- 3 Implantation
- Implantation Procedure
- Having parts ready
- Keeping an external defibrillator ready
- Unpacking the device
- Checking parts
- Implantation site
- Overview: Implanting
- Avoid damage to the header
- Preventing short circuits in the header
- Keeping distance between leads
- Connecting the lead connector to the device
- Applying the programming head
- Establishing wandless telemetry
- Auto-initialization
- Behavior during auto- initialization
- Precautionary Measures while Programming
- Checking the device system
- Performing standard tests and monitoring the patient
- Do not interrupt wandless telemetry during a treatment
- Cancelling telemetry
- Avoiding critical parameter settings
- Manually setting lead polarity
- Setting sensing
- Setting the sensitivity
- Preventing device-induced complications
- Preventing conduction of atrial tachycardia
- Phrenic nerve stimulation that cannot be terminated
- Avoiding risks in the case of exclusive left ventricular pacing
- If an ICD is implanted at the same time, do not permit unipolar pacing
- Recognizing lead failure
- Consider power consumption and service time
- Magnet Response
- Follow-up
- Patient Information
- Replacement Indications
- Explantation and Device Replacement
- Implantation Procedure
- 4 Parameters
- 5 Technical Data
29
Parameters
Timing
Refractory periods
Blanking periods
PMT protection
Parameter Range of values Standard SR DR HF
RV refractory period 200 ... (25) ... 500 ms 250 ms x x x
Atrial refractory
period
AUTO AUTO x x
Atrial refractory
period in the modes
AAI(R); AAT(R); DDT
300 ... (25) ... 775 ms 350 ms x x
LV refractory period 200 ms 200 ms x
AUTO PVARP OFF; ON ON x x
PVARP 175 ... (25) ... 600 ms 225 ms x x
PVARP after PVC PVARP + 150 ms
(max: 600 ms)
Automatically set x x
Parameter Range of values Standard SR DR HF
Far-field protection
after Vs
100 ... (10) ... 220 ms 100 ms x x
Far-field protection
after Vp
100 ... (10) ... 220 ms 150 ms x x
Ventricular blanking
period after Ap
30 ... (5) ... 70 ms 30 ms x x
Parameter Range of values Standard SR DR HF
PMT protection OFF; ON ON x x
VA criterion 250 ... (25) ... 500 ms 350 ms x x