User Manual

Table Of Contents
31
Parameters
Pacing and Sensing
Lead configuration
IEGM recordings
Rates for statistics
Parameter Range of values Standard SR DR HF
Sensing polarity A Unipolar; bipolar Unipolar x x x
Sensing polarity RV Unipolar; bipolar Unipolar x x x
Sensing polarity LV Unipolar; bipolar Unipolar x
Pacing polarity A Unipolar; bipolar Unipolar x x x
Pacing polarity RV Unipolar; bipolar Unipolar x x x
Pacing polarity LV Device type HF:
LV1 tip -> LV2 ring
LV1 tip -> RV ring
LV2 ring -> LV1 tip
LV2 ring -> RV ring
LV1 tip -> housing
LV2 ring -> housing
LV1 tip –>
housing
x
Device type HF QP
LV1 tip -> LV2 ring
LV1 tip -> LV4 ring
LV1 tip -> RV ring
LV1 tip -> housing
LV2 ring -> LV1 tip
LV2 ring -> LV4 ring
LV2 ring -> RV ring
LV2 ring -> housing
LV3 ring -> LV2 ring
LV3 ring -> LV4 ring
LV3 ring -> RV ring
LV4 ring -> LV2 ring
LV4 ring -> RV ring
LV1 tip –>
LV2 ring
x
Parameter Range of values Standard SR DR HF
Number of recordings
(each max. 10 s)
12
20
—xxx
High atrial rate (HAR) OFF; AT; mode switching AT x x x
High ventricular rate
(HVR)
OFF; ON ON x x x
Patient triggering
(triggered by patient)
OFF; ON OFF x x x
Pre-trigger recording 0; 25; 50; 75; 100% 75% x x x
IEGM signal Filtered; Unfiltered Filtered x x x
Parameter Range of values Standard SR DR HF
HAR limit 100 ... (10) ... 250 bpm 200 bpm x x
HVR limit 150 ... (5) ... 200 bpm 180 bpm x x x
HVR counter 4; 8; 12; 16 events 8 events x x x
Start resting period 00:00 ... (1:00 AM) ...
23:00 hh:mm
2:00 hh:mm x x x
Duration of resting
period
0.5 ... (0.5) ... 12 h 4 h x x x
Enable lead check OFF; ON ON x x x