User Manual
Table Of Contents
- Edora 8
- Table of Contents
- 1 Product Description
- 2 General Safety Instructions
- 3 Implantation
- Implantation Procedure
- Having parts ready
- Keeping an external defibrillator ready
- Unpacking the device
- Checking parts
- Implantation site
- Overview: Implanting
- Avoid damage to the header
- Preventing short circuits in the header
- Keeping distance between leads
- Connecting the lead connector to the device
- Applying the programming head
- Establishing wandless telemetry
- Auto-initialization
- Behavior during auto- initialization
- Precautionary Measures while Programming
- Checking the device system
- Performing standard tests and monitoring the patient
- Do not interrupt wandless telemetry during a treatment
- Cancelling telemetry
- Avoiding critical parameter settings
- Manually setting lead polarity
- Setting sensing
- Setting the sensitivity
- Preventing device-induced complications
- Preventing conduction of atrial tachycardia
- Phrenic nerve stimulation that cannot be terminated
- Avoiding risks in the case of exclusive left ventricular pacing
- If an ICD is implanted at the same time, do not permit unipolar pacing
- Recognizing lead failure
- Consider power consumption and service time
- Magnet Response
- Follow-up
- Patient Information
- Replacement Indications
- Explantation and Device Replacement
- Implantation Procedure
- 4 Parameters
- 5 Technical Data
33
Parameters
MRI Program
MRI Program
MRI modes
Modes valid for devices marked ProMRI:
MRI parameters
Preset parameters in the MRI program:
Mode Range of values Standard SR DR HF
MRI program ON; OFF; AUTO OFF x x x
Expiration date Today's date ... (1 day) ...
today's date + 14 days
Today's date
+ 14 days
xxx
MRI mode OFF; A00; V00 Dependent
on
permanent
program
x
OFF; D00; A00; V00 x
OFF; D00; A00; V00;
D00-BiV; V00-BiV
x
Parameter Range of values Standard SR DR HF
Basic rate 70 ... (10) ... 160 bpm 90 bpm x x x
AV delay 110 ms 110 ms x x
VV delay 0 ms 0 ms x
Pulse amplitude A/RV 4.8 V — x x x
Pulse width A/RV 1.0 ms
Pulse amplitude LV 0.2 … (0.2) … 6.0 … (0.5) …
7.5 V
As in
permanent
program
x
Pulse width LV 0.1 … (0.1) … 0.5 … (0.25) …
1.5 ms