User Manual

Table Of Contents
33
Parameters
MRI Program
MRI Program
MRI modes
Modes valid for devices marked ProMRI:
MRI parameters
Preset parameters in the MRI program:
Mode Range of values Standard SR DR HF
MRI program ON; OFF; AUTO OFF x x x
Expiration date Today's date ... (1 day) ...
today's date + 14 days
Today's date
+ 14 days
xxx
MRI mode OFF; A00; V00 Dependent
on
permanent
program
x
OFF; D00; A00; V00 x
OFF; D00; A00; V00;
D00-BiV; V00-BiV
x
Parameter Range of values Standard SR DR HF
Basic rate 70 ... (10) ... 160 bpm 90 bpm x x x
AV delay 110 ms 110 ms x x
VV delay 0 ms 0 ms x
Pulse amplitude A/RV 4.8 V x x x
Pulse width A/RV 1.0 ms
Pulse amplitude LV 0.2 … (0.2) … 6.0 … (0.5) …
7.5 V
As in
permanent
program
x
Pulse width LV 0.1 … (0.1) … 0.5 … (0.25) …
1.5 ms