Parameters
Preset Programs
Preset Programs
Standard and safe program
Mode after auto-initialization:
Lead configuration, determined and set immediately after connection (auto lead
check)
Parameters after auto-initialization:
Parameter Factory
setting
Standard Safe program SR DR HF
Mode VVI VVIR VVI
In the AAI mode, the
safe program is also
AAI.
x
Mode DDD DDDR VVI x x
Parameter Factory
setting
Standard Safe
program
SR DR HF
Pacing polarity A/RV Unipolar Unipolar Unipolar x x x
Pacing polarity LV TCUP TCUP TCUP x
Sensing polarity A/RV Unipolar Unipolar Unipolar x x x
Sensing polarity LV Unipolar Unipolar Unipolar x
Automatic lead check ON ON x x x
Parameter Factory
setting
Standard Safe
program
SR DR HF
Basic rate 60 bpm 60 bpm 70 bpm x x
50 bpm 50 bpm x
Night rate OFF OFF OFF x x x
Rate hysteresis OFF OFF OFF x x x
Upper rate 130 bpm 130 bpm x x
AV dynamics Low Low x x
AV hysteresis mode OFF OFF x x
Sense compensation -45 ms -45 ms x x
AV safety delay 100 ms 100 ms x x
VV delay 0 0 0 x
LV T-wave protection ON ON ON x
Far-field protection after Vs 100 ms 100 ms x x
Far-field protection after Vp 150 ms 150 ms x x
Ventricular blanking period
after Ap
30 ms 30 ms x x
PMT protection ONON— xx
VA criterion 350 ms 350 ms x x
Magnet response AUTO AUTO AUTO x x x
Pulse amplitude A 3.0 V 3.0 V x x
Pulse amplitude RV 3.0 V 3.0 V 4.8 V x x x
Pulse amplitude LV 3.0 V 3.0 V 4.8 V x
Pulse width A 0.4 ms 0.4 ms x x
Pulse width RV 0.4 ms 0.4 ms 1.0 ms x x x
Pulse width LV 0.4 ms 0.4 ms 1.0 ms x
Sensitivity A AUTO AUTO x x
Sensitivity RV AUTO AUTO 2.5 mV x x x
Sensitivity LV AUTO AUTO 2.5 mV x