User Manual
Table Of Contents
- Edora 8
- Table of Contents
- 1 Product Description
- 2 General Safety Instructions
- 3 Implantation
- Implantation Procedure
- Having parts ready
- Keeping an external defibrillator ready
- Unpacking the device
- Checking parts
- Implantation site
- Overview: Implanting
- Avoid damage to the header
- Preventing short circuits in the header
- Keeping distance between leads
- Connecting the lead connector to the device
- Applying the programming head
- Establishing wandless telemetry
- Auto-initialization
- Behavior during auto- initialization
- Precautionary Measures while Programming
- Checking the device system
- Performing standard tests and monitoring the patient
- Do not interrupt wandless telemetry during a treatment
- Cancelling telemetry
- Avoiding critical parameter settings
- Manually setting lead polarity
- Setting sensing
- Setting the sensitivity
- Preventing device-induced complications
- Preventing conduction of atrial tachycardia
- Phrenic nerve stimulation that cannot be terminated
- Avoiding risks in the case of exclusive left ventricular pacing
- If an ICD is implanted at the same time, do not permit unipolar pacing
- Recognizing lead failure
- Consider power consumption and service time
- Magnet Response
- Follow-up
- Patient Information
- Replacement Indications
- Explantation and Device Replacement
- Implantation Procedure
- 4 Parameters
- 5 Technical Data
34
Parameters
Preset Programs
Preset Programs
Standard and safe program
Mode after auto-initialization:
Lead configuration, determined and set immediately after connection (auto lead
check)
Parameters after auto-initialization:
Parameter Factory
setting
Standard Safe program SR DR HF
Mode VVI VVIR VVI
In the AAI mode, the
safe program is also
AAI.
x
Mode DDD DDDR VVI x x
Parameter Factory
setting
Standard Safe
program
SR DR HF
Pacing polarity A/RV Unipolar Unipolar Unipolar x x x
Pacing polarity LV TCUP TCUP TCUP x
Sensing polarity A/RV Unipolar Unipolar Unipolar x x x
Sensing polarity LV Unipolar Unipolar Unipolar x
Automatic lead check ON ON — x x x
Parameter Factory
setting
Standard Safe
program
SR DR HF
Basic rate 60 bpm 60 bpm 70 bpm x x
50 bpm 50 bpm x
Night rate OFF OFF OFF x x x
Rate hysteresis OFF OFF OFF x x x
Upper rate 130 bpm 130 bpm — x x
AV dynamics Low Low — x x
AV hysteresis mode OFF OFF — x x
Sense compensation -45 ms -45 ms — x x
AV safety delay 100 ms 100 ms — x x
VV delay 0 0 0 x
LV T-wave protection ON ON ON x
Far-field protection after Vs 100 ms 100 ms — x x
Far-field protection after Vp 150 ms 150 ms — x x
Ventricular blanking period
after Ap
30 ms 30 ms — x x
PMT protection ONON— xx
VA criterion 350 ms 350 ms — x x
Magnet response AUTO AUTO AUTO x x x
Pulse amplitude A 3.0 V 3.0 V — x x
Pulse amplitude RV 3.0 V 3.0 V 4.8 V x x x
Pulse amplitude LV 3.0 V 3.0 V 4.8 V x
Pulse width A 0.4 ms 0.4 ms — x x
Pulse width RV 0.4 ms 0.4 ms 1.0 ms x x x
Pulse width LV 0.4 ms 0.4 ms 1.0 ms x
Sensitivity A AUTO AUTO — x x
Sensitivity RV AUTO AUTO 2.5 mV x x x
Sensitivity LV AUTO AUTO 2.5 mV x