Technical Data
Battery Data
Battery Data
Battery characteristics
The following data is provided by the manufacturers:
Shortening of the service time
after long storage period
In case of implantation after an average storage period – about 1 year before the
end of the use by date – the average service time decreases by about 1%.
Power consumption
BOS, inhibited: SR(-T), DR(-T) 6 µA; HF-T (QP) 7 µA
BOS, 100% pacing: SR(-T) 8 µA; DR(-T) 11 µA; HF-T (QP) 14 µA
Calculation of service times
Mean service times pre-estimated from the following and other data:
Storage for 6 months
Technical data of the battery manufacturer
Basic rate of 60 bpm in AAIR/VVIR modes (single-chamber devices) or
DDDR modes (dual-chamber and triple-chamber devices)
•Home Monitoring configuration: OFF
No wandless telemetry
Configuration of different pulse amplitudes and lead impedances
Mean service times SR
For single-chamber devices the following times result when set to AAIR or VVIR,
with a basic rate of 60 bpm and a pulse width of 0.4 ms at an impedance of 500 Ω:
Mean service times DR
For dual-chamber devices the following times result when set to DDDR with a basic
rate of 60 bpm and a pulse width of 0.4 ms at an impedance of 500 Ω:
Manufacturer Wilson GREATBATCH,
INC. Clarence, NY 14031
LITRONIK GmbH,
01796 Pirna, Germany
Battery type GB 3193 LiS 2650MK LiS 3150MK
System QMR LiMn0
2
LiMn0
2
Device type SR; DR HF; HF QP
Battery voltage at BOS 3.3 V 3.1 V 3.1 V
Open-circuit voltage 3.3 V 3.1 V 3.1 V
Nominal capacity 1010 mAh 950 mAh 1200 mAh
Usable capacity until EOS 971 mAh 880 mAh 1066 Ah
Remaining capacity at ERI 39 mAh 70 mAh 134 mAh
Amplitude Pacing Average service time
2.5 V 100% 13 years
50% 14 years, 9 months
3.0 V 100% 11 years, 3 months
50% 13 years, 7 months
5.0 V 100% 5 years, 6 months
Amplitude Pacing Average service time
A: 2.5 V
RV: 2.5 V
100% 9 years, 4 months
50% 11 years, 4 months
A: 3.0 V
RV: 3.0 V
100% 7 years, 8 months
50% 10 years
A: 5.0 V
RV: 5.0 V
100% 3 years, 2 months