User Manual

Table Of Contents
en • English
23
Tachycardia
Detection
Therapy: Atrial therapy
The following parameters apply to devices of series 7:
Therapy: Ventricular ATP
Parameter Range of values Standard VR DX DR HF
Interval AT/AF 240 ... 600 ms 300 ms x x x
Interval VT1 OFF; 270 ... (10) ... 600 ms OFF xxxx
Interval VT2 OFF; 270 ... (10) ... 500 ms
Interval VF OFF; 240 ... (10) ... 400 ms 300 ms
Detection counter VT1 10 ... (2) ... 100 28 xxxx
Detection counter VT2 10 ... (2) ... 80 20
Detection counter VF 6 out-of 8; 8 out-of 12;
10 out-of 14; 12 out-of 16;
16 out-of 20; 18 out-of 24;
20 out-of 26; 22 out-of 30;
24 out-of 30; 30 out-of 40
18 out of
24
Redetection counter VT1 10 ... (2) ... 50 20 xxxx
Redetection counter VT2 10 ... (2) ... 40 14
Redetection counter VF 6 out-of 8; 8 out-of 12;
10 out-of 14; 12 out-of 16;
16 out-of 20;
18 out-of 24; 20 out-of 26;
22 out-of 30; 24 out-of 30
8 out of
12
xxxx
SMART detection VT1/VT2 OFF; ON ON x x x
SMART detection ON:
– Onset VT1/VT2 4 ... (4) ... 32% 20% x x x
– Stability VT1/VT2 8 ... (4) ... 48% 12%
SMART detection OFF:
– Onset VT1/VT2 OFF; 4 ... (4) ... 32% 20% xxxx
– Stability VT1/VT2 OFF; 8 ... (4) ... 48 ms
8 ... (4) ... 48%
12%
MorphMatch OFF; Monitoring; ON OFF xxxx
MorphMatch threshold Low; Std; High Std. xxxx
Sustained VT OFF; 1; 2; 3; 5; 10; 20;
30 min
OFF xxxx
Parameter Range of values Standard VR DX DR HF
Atrial therapy in the presence of stable atrial flutter:
ATP type OFF; Burst; Ramp OFF x x
Number S1 2 ... (1) ... 10 5 x x
P-S1 interval 70 ... (5) ... 95% 80% x x
S1 decrement 5 ... (5) ... 40 ms 10 ms x x
Backup stimulation OFF; 70; 90; OFF x x
Atrial therapy in the presence of unstable atrial fibrillation:
Therapy OFF; HF (high frequency)
burst
OFF x x
Rate 10 ... (5) ... 40 Hz 40 Hz x x
Duration 2 ... (1) ... 10 3 s x x
Backup stimulation OFF; 70; 90; OFF x x
Parameter Range of values Standard VR DX DR HF
For VT1/VT2: Attempts OFF; 1 ... (1) ... 10 OFF xxxx
ATP type for VT1/VT2 Burst; Ramp Burst xxxx
ATP type for VF OFF; Burst; Ramp Burst xxxx
ATP optimization OFF; ON OFF xxxx
Number S1 for VT1/VT2 1 ... (1) ... 15 5 xxxx
Number S1 for VF 8
S1 decrement for VT1/VT2
and for VF
5 ... (5) ... 40 ms 10 ms xxxx
Scan decrement for
VT1/VT2
OFF; 5 ... (5) ... 40 ms OFF xxxx
Add S1 for VT1/VT2 OFF; ON ON xxxx
Ventricular pacing RV; LV; BiV RV x
417634--B_GA_Intica-ProMRI_mul.fm Page 23 Friday, November 6, 2015 8:13 PM