User Manual

Table Of Contents
24
Therapy: Shock
Sensing
Sensitivity and thresholds
R-S1 interval for VT1/VT2 70 ... (5) ... 85; 88; 90; 95% 80% xxxx
R-S1 interval for VF 88%
Early ATP delivery for VFOFF; ON OFF xxxx
Parameter Range of values Standard VR DX DR HF
Number of shocks VT1/VT20; 1; 2; 6; 8 8 xxxx
Number of shocks VF 6; 8 8 xxxx
1st Shock for VT1/VT2 OFF; 2 ... (2) ... 20 ... (5)
... 40 J
40 J xxxx
2nd Shock for VT1/VT2 OFF; 4 ... (2) ... 20 ... (5)
... 40 J
40 J xxxx
3rd - nth shock for
VT1/VT2
OFF; 4*40 J; 6*40 J 6*40 J xxxx
1st Shock for VF 2 ... (2) ... 20 ... (5) ... 40 J 40 J xxxx
2nd Shock for VF 4 ... (2) ... 20 ... (5) ... 40 J 40 J xxxx
3rd - nth shock for VF 4*40 J; 6*40 J 6*40 J xxxx
For shock in VT1/VT2 and VF:
Confirmation OFF; ON ON xxxx
– Polarity Normal; inverse;
Normal -> alternating;
Inverse -> alternating
Normal
– Shock form Biphasic; Biphasic 2;
Biphasic -> alternating;
biphasic 2 -> alternating
Biphasic
– Shock path RV -> housing + SVC
RV -> housing
RV -> SVC
RV->
ICD+SVC
xxx
RV -> ICD x
Parameter Range of values Standard VR DX DR HF
Parameter Range of values Standard VR DX DR HF
Sensing A STD; OFF STD x x x
Sensing RV STD; TWS; VFS; IND STD xxxx
Sensing LV STD; OFF; IND STD x
DX sensing ON; OFF OFF x
Upper threshold RV 50; 75% 50% xxxx
Upper threshold LV 50; 75% 50% x
Upper threshold
duration RV after detection
110; 150 ... (50) ... 500 ms
VFS: 110 ms
350 ms xxxx
Upper threshold
duration RV after pace
400 ms
Lower threshold RV 25; 50% 25% xxxx
T-wave suppression
after pacing
OFF; ON OFF xxxx
Minimum threshold A 0.2 ... (0.1) ... 2.0 mV 0.4 mV x x x
Minimum threshold RV 0.5 ... (0.1) ... 2.5 mV 0.8 mV xxxx
Minimum threshold LV 0.5 ... (0.1) ... 2.5 ... (0.5)
... 5.0 mV
1.6 mV x
417634--B_GA_Intica-ProMRI_mul.fm Page 24 Friday, November 6, 2015 8:13 PM