User Manual
Table Of Contents
- Cover
- 1 Product Description
- 2 General Safety Instructions
- 3 Implantation
- Implantation Procedure
- Having parts ready
- Keeping an external defibrillator ready
- Unpacking the device
- Checking parts
- Implantation site
- Preventing leakage currents
- Preventing unintentional shock delivery
- Avoiding damage to the header
- Preventing short circuits in the header
- Ensure that connector ports are clean
- Overview: Implanting
- Connecting the device
- Keeping distance between leads
- Applying the programming head
- Establishing RF telemetry
- Activating ICD therapy
- Precautionary Measures while Programming
- Performing standard tests and monitoring the patient
- Cancelling telemetry
- Avoiding critical parameter settings
- Avoiding risks in the case of exclusive LV pacing
- Monitoring the patient when setting asynchronous modes
- Complying with the morphology criteria
- Setting sensing
- Preventing device-induced complications
- Preventing conduction of atrial tachycardia
- Avoiding AV crosstalk
- Observing the shock impedance limit
- Preventing recurrence after therapy shock
- Phrenic nerve stimulation that cannot be terminated
- Note the reduced pulse amplitude due to a battery voltage drop
- Observe when inducing short-term cardiac arrest
- Checking the settings of the DX lead
- Checking for electrodes suitable for the shock path
- Recognizing lead failure
- Considering power consumption and service time
- Magnet Response
- Follow-up
- Patient Information
- Replacement Indications
- Explantation and Device Replacement
- Implantation Procedure
- 4 Parameters
- Bradycardia / CRT
- General ICD therapy
- Timing: Basic rate day/night and rate hystereses
- Timing: AV delay
- Timing: Post-shock pacing
- Timing: Upper rate
- Timing: Mode switching
- Timing: Ventricular pacing suppression
- Timing: Ventricular pacing
- Timing: Ventricular multi pole pacing
- Timing: Refractory periods and blanking periods
- Timing: PMT protection
- Timing: Rate adaptation via accelerometer
- Timing: Rate adaptation via CLS
- Pacing: Pulse amplitude and pulse width
- Pacing: Atrial capture control
- Pacing: Ventricular capture control
- Lead configuration LV on IS-1 connection
- Lead configuration LV on IS4 connection
- MRI program
- Tachycardia
- Sensing
- Diagnostics
- Home Monitoring
- Bradycardia / CRT
- 5 Technical Data
en • English
25
Diagnostics
The following can be set:
Home Monitoring
Parameter Range of values Standard VR DX DR HF
For AT/AF OFF; ON
7 series: Extended ON
ON xxx
For SVT OFF; ON ON x x x
For nsVT OFF; ON ON xxxx
Periodic recording When Home Monitoring
is deactivated:
OFF; 30 ... (30) ...
180 days
90 days xxxx
IEGM configuration RA, RV, LV
RA, RV, FF
FF; RV; LV
RA, RV, LV x
Start resting period 00:00 ... (1:00 AM) ...
23:00 hh:mm
2:00 AM
hh:mm
xxxx
Duration of resting
period
0.5 ... (0.5) ... 12 h 4 h xxxx
AV delay adjusted in
sensing test
OFF; 300 ms 300 ms x x x
Thoracic impedance (TI)OFF, ON OFF xxxx
Parameter Range of values Standard VR DX DR HF
Home Monitoring OFF; ON OFF xxxx
Time of transmission STD; 00:00 ... (1:00 AM)
... 23:00 hh:mm
STD xxxx
IEGM for therapy
episodes
OFF; ON ON xxxx
IEGM for monitoring
episodes
Ongoing atrial episode OFF; 6; 12; 18 h 12 h x x x
Configurable in the HMSC:
Transmission date XX.XX.XXXX Follow-up
+ 91 days
xxxx
Cycle duration 20 ... (1) ... 366 days 91 days xxxx
417634--B_GA_Intica-ProMRI_mul.fm Page 25 Friday, November 6, 2015 8:13 PM