User Manual
Table Of Contents
- Cover
- 1 Product Description
- 2 General Safety Instructions
- 3 Implantation
- Implantation Procedure
- Having parts ready
- Keeping an external defibrillator ready
- Unpacking the device
- Checking parts
- Implantation site
- Preventing leakage currents
- Preventing unintentional shock delivery
- Avoiding damage to the header
- Preventing short circuits in the header
- Ensure that connector ports are clean
- Overview: Implanting
- Connecting the device
- Keeping distance between leads
- Applying the programming head
- Establishing RF telemetry
- Activating ICD therapy
- Precautionary Measures while Programming
- Performing standard tests and monitoring the patient
- Cancelling telemetry
- Avoiding critical parameter settings
- Avoiding risks in the case of exclusive LV pacing
- Monitoring the patient when setting asynchronous modes
- Complying with the morphology criteria
- Setting sensing
- Preventing device-induced complications
- Preventing conduction of atrial tachycardia
- Avoiding AV crosstalk
- Observing the shock impedance limit
- Preventing recurrence after therapy shock
- Phrenic nerve stimulation that cannot be terminated
- Note the reduced pulse amplitude due to a battery voltage drop
- Observe when inducing short-term cardiac arrest
- Checking the settings of the DX lead
- Checking for electrodes suitable for the shock path
- Recognizing lead failure
- Considering power consumption and service time
- Magnet Response
- Follow-up
- Patient Information
- Replacement Indications
- Explantation and Device Replacement
- Implantation Procedure
- 4 Parameters
- Bradycardia / CRT
- General ICD therapy
- Timing: Basic rate day/night and rate hystereses
- Timing: AV delay
- Timing: Post-shock pacing
- Timing: Upper rate
- Timing: Mode switching
- Timing: Ventricular pacing suppression
- Timing: Ventricular pacing
- Timing: Ventricular multi pole pacing
- Timing: Refractory periods and blanking periods
- Timing: PMT protection
- Timing: Rate adaptation via accelerometer
- Timing: Rate adaptation via CLS
- Pacing: Pulse amplitude and pulse width
- Pacing: Atrial capture control
- Pacing: Ventricular capture control
- Lead configuration LV on IS-1 connection
- Lead configuration LV on IS4 connection
- MRI program
- Tachycardia
- Sensing
- Diagnostics
- Home Monitoring
- Bradycardia / CRT
- 5 Technical Data
en • English
29
Intica 5/7 VR-T
Service times with GB 2992 or LiS 3410 RR battery:
Intica 5/7 VR-T DX
Service times with GB 2992 or LiS 3410 RR battery:
Intica 5/7 DR-T
Service times with GB 2992 or LiS 3410 RR battery:
Intica 5 HF-T (QP)
Service times with GB 2992 or LiS 3410 RR battery:
Intica 7 HF-T (QP)
Service times with battery GB 2992 without multipolar pacing:
Service times with battery GB 2992 with multipolar pacing:
Pacing
Service time [in years] at number of shocks per year
48121620
0%
10.3 8.3 7.0 6.0 5.3
15%
10.1 8.1 6.8 5.9 5.2
50%
9.5 7.8 6.6 5.7 5.0
100%
8.8 7.3 6.2 5.4 4.8
Pacing
Service time [in years] at number of shocks per year
48121620
0%
9.4 7.7 6.5 5.7 5.0
15%
9.2 7.6 6.4 5.6 4.9
50%
8.7 7.2 6.2 5.4 4.8
100%
8.1 6.8 5.9 5.2 4.6
Pacing
Service time [in years] at number of shocks per year
48121620
0%
9.4 7.7 6.5 5.7 5.0
15%
9.0 7.4 6.3 5.5 4.9
50%
8.1 6.8 5.9 5.2 4.6
100%
7.1 6.1 5.3 4.7 4.3
Pacing
Service time [in years] at number of shocks per year
4 8 121620
0%
8.9 7.4 6.3 5.5 4.9
15%
8.3 7.0 6.0 5.2 4.7
50%
7.2 6.1 5.4 4.8 4.3
100%
6.0 5.3 4.7 4.2 3.9
Pacing
Service time [in years] at number of shocks per year
4 8 121620
0%
10.1 8.4 7.2 6.3 5.5
15%
9.4 7.9 6.8 6.0 5.3
50%
8.2 7.0 6.1 5.5 4.9
100%
6.9 6.0 5.4 4.8 4.4
Pacing
Service time [in years] at number of shocks per year
4 8 121620
0%
10.1 8.4 7.2 6.3 5.5
15%
9.2 7.8 6.7 5.9 5.3
50%
7.7 6.6 5.9 5.2 4.7
100%
6.2 5.5 5.0 4.5 4.1