User Manual
Table Of Contents
- Cover
- 1 Product Description
- 2 General Safety Instructions
- 3 Implantation
- Implantation Procedure
- Having parts ready
- Keeping an external defibrillator ready
- Unpacking the device
- Checking parts
- Implantation site
- Preventing leakage currents
- Preventing unintentional shock delivery
- Avoiding damage to the header
- Preventing short circuits in the header
- Ensure that connector ports are clean
- Overview: Implanting
- Connecting the device
- Keeping distance between leads
- Applying the programming head
- Establishing RF telemetry
- Activating ICD therapy
- Precautionary Measures while Programming
- Performing standard tests and monitoring the patient
- Cancelling telemetry
- Avoiding critical parameter settings
- Avoiding risks in the case of exclusive LV pacing
- Monitoring the patient when setting asynchronous modes
- Complying with the morphology criteria
- Setting sensing
- Preventing device-induced complications
- Preventing conduction of atrial tachycardia
- Avoiding AV crosstalk
- Observing the shock impedance limit
- Preventing recurrence after therapy shock
- Phrenic nerve stimulation that cannot be terminated
- Note the reduced pulse amplitude due to a battery voltage drop
- Observe when inducing short-term cardiac arrest
- Checking the settings of the DX lead
- Checking for electrodes suitable for the shock path
- Recognizing lead failure
- Considering power consumption and service time
- Magnet Response
- Follow-up
- Patient Information
- Replacement Indications
- Explantation and Device Replacement
- Implantation Procedure
- 4 Parameters
- Bradycardia / CRT
- General ICD therapy
- Timing: Basic rate day/night and rate hystereses
- Timing: AV delay
- Timing: Post-shock pacing
- Timing: Upper rate
- Timing: Mode switching
- Timing: Ventricular pacing suppression
- Timing: Ventricular pacing
- Timing: Ventricular multi pole pacing
- Timing: Refractory periods and blanking periods
- Timing: PMT protection
- Timing: Rate adaptation via accelerometer
- Timing: Rate adaptation via CLS
- Pacing: Pulse amplitude and pulse width
- Pacing: Atrial capture control
- Pacing: Ventricular capture control
- Lead configuration LV on IS-1 connection
- Lead configuration LV on IS4 connection
- MRI program
- Tachycardia
- Sensing
- Diagnostics
- Home Monitoring
- Bradycardia / CRT
- 5 Technical Data
Note:
Not all functions and parameters mentioned in this technical manual are
featured by each device type of each device family.
en • English
5
Modes
The mode setting depends on the individual diagnosis:
NBD and NBG codes
VVE is the NBD code for the antitachycardia mode of the single-chamber, dual-
chamber, and triple-chamber devices without atrial therapy:
VDE is the NBD code for the antitachycardia mode of the dual-chamber and triple-
chamber devices with atrial therapy:
DDDR is the NBG code for the antibradycardia mode of the dual-chamber devices:
DDDRV is the NBG code for the antibradycardia mode of the triple-chamber devices:
VDDR is the NBG code for the antibradycardia mode of the single-chamber type DX
device:
VVIR is the NBG code for the antibradycardia pacing modes of the single-chamber
device:
Device type Modes
VR VVI; VVIR; VOO; OFF
7 series: VVI-CLS
VR DX VDD; VDDR; VDI; VDIR; VVI; VVIR; V00; OFF
7 series: VVI-CLS
DR, HF (QP) DDD; DDDR; DDI; DDIR
DDDR-ADIR; DDD-ADI
VDD; VDDR; VDI; VDIR
VVI; VVIR; AAI; AAIR; VOO; DOO; OFF
7 series: VVI-CLS; DDD-CLS
V Shock in the ventricle
V Antitachycardia pacing (ATP) in the ventricle
E Detection via IEGM analysis
V Shock in the ventricle
D Antitachycardia pacing (ATP) in the atrium and ventricle
E Detection via IEGM analysis
D Pacing in the atrium and ventricle
D Sensing in the atrium and ventricle
D Pulse inhibition and pulse triggering
R Rate adaptation
D Pacing in the atrium and ventricle
D Sensing in the atrium and ventricle
D Pulse inhibition and pulse triggering
R Rate adaptation
V Multisite pacing in both ventricles
V Ventricular pacing
D Sensing in the atrium and ventricle
D Pulse inhibition and pulse triggering
R Rate adaptation
V Ventricular pacing
V Sensing in the ventricle
I Pulse inhibition in the ventricle
R Rate adaptation