User Manual

Table Of Contents
14 General Safety Instructions
Possible Complications
General information on
medical complications
Complications for patients and device systems generally recognized among prac-
titioners also apply to BIOTRONIK devices.
Normal complications may include fluid accumulation within the device pocket,
infections, or tissue reactions. Primary sources of complication information
include current scientific and technological knowledge.
It is impossible to guarantee the efficacy of antitachycardia therapy, even if the
programs have proven successful during tests or subsequent electrophysio-
logical examinations. In rare cases the set parameters may become ineffective.
It is possible for therapies to induce or accelerate tachycardia and cause sus-
tained ventricular flutter or fibrillation.
Skeletal myopotentials Bipolar sensing and control of sensitivity are adapted by the device to the rate
spectrum of intrinsic events so that skeletal myopotentials are usually not
recorded. Skeletal myopotentials can nonetheless be classified as intrinsic events
especially at very high sensing sensitivity and, depending on the interference, may
cause inhibition or antiarrhythmia therapy.
In the case of undesired myopotentials, the device switches to asynchronous
pacing if the interference rate is exceeded.
Possible technical failures Technical failure of a device system cannot be entirely ruled out. Possible causes
can include the following:
Lead dislodgement, lead fracture
Insulation defects
Device component failures
Battery depletion
Interrupted telemetry
Electromagnetic interfer-
ence (EMI)
Any device can be sensitive to interference if external signals are sensed as
intrinsic rhythm or if measurements prevent rate adaptation.
BIOTRONIK devices have been designed so that their susceptibility to EMI is
minimal.
Due to the intensity and variety of EMI, there is no guarantee for safety. It is
generally assumed that EMI produces only minor symptoms, if any, in patients.
Depending on the pacing mode and the type of interference, sources of inter-
ference may lead to pulse inhibition or triggering, an increase in the sensor-
dependent pacing rate or asynchronous pacing.
Under unfavorable conditions, for example during therapeutic or diagnostic
procedures, interference sources may induce such a high level of energy into
the pacing system that the cardiac tissue surrounding the lead tip is damaged.
Device behavior in case of
EMI
In case of electromagnetic interference, the device switches to asynchronous
pacing for as long as the interference rate is exceeded.
Static magnetic fields The reed switch in the device closes starting at a field strength of 1.8 mT. The reed
switch opens if the magnetic field falls below 1 mT.