User Manual

Table Of Contents
19 Implantation
Connecting the lead
connector to the device
Keeping distance between
leads
1 Disconnect stylets and stylet guides.
2 DF-1/IS-1 connection:
Connect the DF-1 connector for the right-ventricular shock coil to RV.
Connect the DF-1 connector for the supraventricular shock coil to SVC.
Or connect a subcutaneous array to SVC.
DF4/IS-1 connection:
Connect the DF4 connector to RV
3 DF-1/IS-1 connection:
Connect the bipolar IS-1 lead connector for the atrium to RA.
Connect the IS-1 lead connector for the right ventricle to RV.
Connect the unipolar or the bipolar IS-1 lead connector for the left ven-
tricle to LV.
DF4/IS-1 connection:
Connect the bipolar IS-1 lead connector for the atrium to RA.
Connect the unipolar or the bipolar IS-1 lead connector for the left ven-
tricle to LV.
4 Push the lead connector into the header without twisting or bending the
connector or conductor until the connector tip (on the DF-1 connector) or
the insertion indicator (on the DF4 connector) becomes visible behind the
set screw block. This indicator can vary depending on the manufacturer of
the lead used.
5 If you cannot easily plug the lead connector into the connection:
Use only sterile water as lubricant.
6 If the lead connector cannot be inserted completely, the set screw may be
protruding into the drill hole of the set screw block.
Use the screwdriver to perpendicularly pierce through the slitted point
in the center of the silicone plug until it reaches the set screw.
Carefully loosen the set screw without completely unscrewing it, so that
it does not become tilted upon retightening.
7 Turn the set screw clockwise until torque control starts (you will hear a
clicking sound).
8 Carefully withdraw the screwdriver without retracting the set screw.
In case of IS-1 connections with two set screws, tighten both screws!
When the screwdriver is withdrawn, the silicone plug automatically
safely seals the lead connector port.
!
!
WARNING
Inadequate therapy
When leads are not spaced sufficiently apart or are positioned inappropriately,
this can lead to far-field sensing or insufficient defibrillation.
The distance between 2 shock coils must be greater than 6 cm.
Tip and ring electrodes must not have contact with each other.