User Manual
Table Of Contents
- Cover
- Table of Contents
- 1 Product Description
- 2 General Safety Instructions
- 3 Implantation
- Implantation Procedure
- Having parts ready
- Keeping an external defibrillator ready
- Unpacking the device
- Checking parts
- Implantation site
- Preventing leakage currents
- Preventing unintentional shock delivery
- Avoiding damage to the header
- Preventing short circuits in the header
- Ensure that connections are clean
- Connecting the lead connector to the device
- Keeping distance between leads
- Implanting
- Applying the programming head
- Establishing telemetry contact
- Activating ICD therapy
- Precautionary Measures while Programming
- Performing standard tests and monitoring the patient
- Cancelling telemetry
- Avoiding critical parameter settings
- Check for leads suitable for shock path
- Monitoring the patient when setting asynchronous modes
- Setting sensing
- Preventing device-induced complications
- Preventing conduction of atrial tachycardia
- Observing the shock impedance limit
- Preventing recurrence after therapy shock
- Phrenic nerve stimulation that cannot be terminated
- Avoiding risks in the case of exclusive LV pacing
- Recognizing lead failure
- Considering power consumption and service time
- Magnet Response
- Follow-up
- Patient Information
- Replacement Indications
- Explantation and Device Replacement
- Implantation Procedure
- 4 Parameters
- Bradycardia / CRT
- General ICD therapy
- Timing: Basic rate day/ night and rate hystereses
- Timing: AV delay
- Timing: Post-shock pacing
- Timing: Upper rate
- Timing: Mode switching
- Timing: Ventricular pacing
- Timing: Refractory periods and blanking periods
- Timing: PMT protection
- Timing: Rate adaptation via accelerometer
- Pacing: Pulse amplitude and pulse width
- Pacing: Ventricular capture control
- Pacing: atrial capture control
- LV lead configuration
- Tachycardia
- Sensing
- Diagnostics
- Home Monitoring
- Bradycardia / CRT
- 5 Technical Data
29 Parameters
4 Parameters
Parameters4GA-HW_en--mul_393468-BTechnical[nbsp ]Manual for the[nbsp ]DeviceIlesto 5/7 V R- T, VR-T DX, DR-T, HF-T
Bradycardia / CRT
General ICD therapy
Timing: Basic rate day/
night and rate hystereses
Parameter Range of values Standard
VR
DX
DR
HF
ICD therapy OFF; ON ON xxxx
Programs Display standard program; Display
safe program; Display first interro-
gated program; Individual 1,2,3
– xxxx
Parameter Range of values Standard
VR
DX
DR
HF
Basic rate 30 ... (5) ... 100 ... (10) ... 160 bpm 40 bpm x x
60 bpm x x
Night rate OFF; 30 ... (5) ... 100bpm OFF xxxx
Begin of night 00:00 ... (00:01) ...23:59hh:mm 06:00 hh:mmxxxx
End of night 22:00 hh:mm
Rate hysteresis OFF
-5 ... (-5) ... -25 ... (-20) ... -65 bpm
OFF xxxx
Scan/repetitive OFF; ON ON xxxx