User Manual

Table Of Contents
41 Technical Data
Resistance to interference Note on device type DX (only devices with a DF-1/IS-1 connection): The EMC
requirements are met as long as atrial sensitivity is set to 1.0 mV (factory set-
tings) or values 1.0 mV. Measures must be taken to assure interference-free
therapy if more sensitive values are set.
Note on device type HF: In the case of unipolar sensing, the requirement for
interference voltages of ≤ 0.3 mV (tip to tip) is met.
Common mode rejection
ATP amplitude A burst was measured at 500 Ω, an amplitude of 7.5 V (tolerance ±1.5 V), pulse
width of 1.5 ms, R-S1 interval of 300 ms and an S1 count of 5:
Automatic sensitivity
setting
Measurement of actual values and test signal wave shape: standard triangle. For
the device type DX, the programmed atrial sensitivity is intensified by a factor of 4.
Shock energy / peak voltage With shock path: RV to housing + SVC
Rate Common mode rejection ratio
Atrium: DX* Atrium: DR,
HF
V right: VR, DR,
HF
V left: HF
16.6 Hz 58 dB 53 dB 64 dB 66 dB
50 Hz 55 dB 55 dB 64 dB 66 dB
60 Hz 56 dB 56 dB 64 dB 68 dB
* only devices with a DF-1/IS-1 connection.
ATP amplitude Measured
minimum
Measured
maximum
Mean value
RV 7.67 V 7.67 V 5.00 V
LV 7.67 V 7.67 V 4.99 V
Sensitivity Value Tolerance Measured value
A: positive 0.2 mV 0.2 ... 0.5 0.24 mV
A: negative 0.24 mV
DX: A: positive 0.2 mV 0.2 ... 0.52
(0.05 to 0.13)
0.05 mV
DX:A:negative 0.05mV
RV: positive 0.5 mV 0.3 ... 0.7 0.48 mV
RV: negative 0.40 mV
LV: positive 0.5 mV 0.3 ... 0.7 0.48 mV
LV: negative 0.56 mV
Shock energy (Toler-
ance)
Tolerance peak
voltage
Measured
value
Shock energy
Measured
value
Peak voltage
1 J (0.7 ... 1.18) 90 ... 120 V 0.84 J 100 V
20 J (16.9 ... 20.9) 440 ... 480 V 18.1 J 469 V
40 J (33.8 ... 41.4) 620 ... 690 V 36.9 J 667 V