Warranty

BioWaveGO User’s Manual
BioWave Corporation84
1. Lotion on the skin.
Lotion dramatically reduces adhesion and conductivity of the Pain Relief Pads and
can cause this error condition to appear. Use a washcloth with soap and water to
clean the skin well and then dry thoroughly. Use a new set of Pain Relief Pads and
place onto the cleaned skin. The 5 LEDs should stop blinking ORANGE and the
LED bar graph will change to show remaining treatment time in WHITE. You may
then start the treatment by pressing the PLUS button to ramp up the intensity to a
strong but comfortable level.
2. The Pain Relief Pad has been used too many times and has lost its
tackiness and adhesion.
Use a new set of Pain Relief Pads to establish proper electrical contact with the
skin.
3. Dryakyskin.
Dryakyskinorheavilysuntannedskincancausethestimulatornottorecognize
the Pain Relief Pads because the impedance of the skin is too high. Remove the
PainReliefPadsfromtheskin.Iftheyhaveasignicantconcentrationofwhiteecks
(dead skin) stuck to the surface then discard them. Use a washcloth with soap and
water to clean the skin well and then dry thoroughly. Use a new set of Pain Relief
Pads and place onto the cleaned skin. The 5 LEDs should stop blinking ORANGE
and the LED bar graph should change to show remaining treatment time in
WHITE. You may then start the treatment by pressing the PLUS button to ramp up
the intensity to a strong but comfortable level.
4. Excessive hair on skin.
Excessive hair on the skin can prevent the Pain Relief Pads from having adequate
adhesion and electrical contact with the skin. As a result, the stimulator may not