Warranty

BioWave Corporation 85
8. Troubleshooting
recognize the Pain Relief Pads even though they appear to be properly placed
on the skin. Remove the Pain Relief Pads from the skin and place them onto the
plastic liner from which they came. Use a razor to shave the area where each Pain
Relief Pad is to be placed. Place the Pain Relief Pads back into position on the
clean shaved skin. The 5 LEDs should stop blinking ORANGE and the LED bar
graph will change to show remaining treatment time in WHITE. You may then start
the treatment by pressing the PLUS button to ramp up the intensity to a strong
but comfortable level.
5. Pain Relief Pad(s) pull off of the skin.
If during the procedure, either Pain Relief Pad or both Pain Relief Pads get pulled
off of your skin, the intensity will go to zero (0.0%), the treatment will pause and
the 5 LEDs will blink ORANGE.
Make sure the Pain Relief Pads are clean and have not picked up any debris on
the hydrogel. If they are clean, place the Pain Relief Pads back onto your skin in
the correct location and the 5 LEDs should stop blinking ORANGE and the LED
bar graph will change to show remaining treatment time in WHITE. If the LED Bar
Graph is still blinking ORANGE, place a new set of Pain Relief Pads on yourself in
the correct location. The 5 LEDs should stop blinking ORANGE and the LED bar
graph will change to show remaining treatment time in WHITE. You then have
to manually ramp up the intensity from zero back to the prior treatment level to
continue the treatment.
8.2 Use of Non-BioWave Pain Relief Pads
BioWave Pain Relief Pads must be used. The BioWaveGO device will only recognize
and work with a BioWave Leadwire Cable and BioWave Pain Relief Pads.