Warranty

BioWaveGO User’s Manual
BioWave Corporation92
The BioWaveGO Neurostimulator is intended for use in the electromagnetic environment specified below. The
customer or the user of the BioWaveGO Neurostimulator should assure that it is used in such an environment.
±6 kV contact
±8 kV air
± 2 kV for power
± 1 kV for input/
output lines
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
<5% U
T
(>95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70 % U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
3 A/m
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage
variations on power
supply input lines
IEC 61000-4-11
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic
environment – guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency
(50/60 Hz)
IEC 61000-4-8
±1 kV differential mode
±2 kV common mode
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
Earth
Mains power quality should be
that of a typical commercial or
hospital environment.
<5% U
T
(>95% dip
in U
T
) for 0,5 cycle
40% U
T
(60% dip
in U
T
) for 5 cycles
70% U
T
(30% dip
in U
T
) for 25 cycles
<5% U
T
(>95% dip
in U
T
for 5s)
Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the BioWaveGO requires continued
operation during power mains
interruptions, it is recommended
that the BioWaveGO be powered
from an uninterruptible power
supply or a battery.
3 A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
NOTE: U
T
is the A.C. mains voltage prior to application of the test level.
Guidance & Manufacturer’s Declaration – Electromagnetic Immunity