Manual
12
Uncertainty
b
± 8% of full-scale peak 
envelope power (95% 
c.l.)
Match Measurement
Match Range
Return Loss
Rho (
ρ)
VSWR
0 to 20 dB
0.1 to 1
1.22 to 99.99
Uncertainty Twice the Avg Power 
Uncertainty
(Calculated from 
forward and reflected 
uncertainty)
Settling Time, Max 2.5 seconds
Impedance, Nominal 50 ohms
Insertion Loss, Max 0.05 dB up to 1 GHz
Input VSWR, Max. 1.05:1 up to 1 GHz
Directivity, Typicala 30 dB
RF Connectors QC Type (N(F) normally 
supplied)
Power Supply From host instrument 
via cable
Mechanical Shock and Vibration In accordance with MIL-
PRF-28800F Class 3
CE CE compliant. Refer to 
DOC for specific 
standards.
Recommended Calibration 
Interval
1 year