User Manual
8
Table 3 has two examples of uncertainty calculations. 
The first is a 10 mW signal at room temperature. The 
second is a 10 µW, 40 MHz signal at 50°C. Since this 
measurement is at both low frequency and high tem
-
perature, the uncertainty will be increased. Note that 
the RSS uncertainties are smaller than the values 
from the rough estimate. This will always be the case.
Table 3 - Uncertainty Examples
Example 1
(10 mW, 
Room Temp)
Example 2
(10 µW, 
40 MHz, 50°C)
Error 
Source
Percent 
Uncert.
RSS 
Term
Percent 
Uncert.
RSS 
Term
Cal. Uncert. 1.13 %   1.28 1.13 %   1.28
Freq. Resp. 3.42 % 11.70 3.42 % 11.70
Temp. Lin. 3.43 % 11.76 3.43 % 11.76
Other 0.5 %   0.25 1.6 %   2.56
Res. 0.005 %  0.00 0.5 %  0.25
Zero Set 0.00125 %   0.00 1.25 %   1.56
Noise 0.007 %  0.00 2 %  4.00
Sum Uncert. 24.99 33.11
RSS Uncert.  5.00 %  5.75 %
Quick Uncert.   5.01 % 16 %