Owner manual

1
Introduction
Connecting a mismatch of a known value and measuring it with the Antenna Tes-
ter provides a test to determine the degree of measurement accuracy.
Measurement accuracy is specified to be ± 0.040 of Rho (p), where:
Because of the manufacturing tolerance, each mismatch is analyzed across the
frequency band of the Antenna Tester. Rho data is recorded so that future com-
parisons can be made without regard to the manufacturing tolerance.
This verification kit is used to verify Antenna Tester performance. It is intended
for use with the AT-800 Antenna Tester. These instructions were written with the
understanding that the operator is familiarwith the operation of the Antenna Tes-
ter. Refer to the Antenna Tester Operator Manualfor more information.
Kit Contents and Specifications
Standard Mismatch—p/n 2-T-MN-1R50 (includes verification data)
VSWR 1.5:1 (nominal)
Rho () 0.200 (nominal)
Connector Type N (M)
Power Rating 1 Watt cw maximum
Operating Temperature 20° to 30°C (68° to 86°F)
Storage Temperature 0° to 50°C (32° to 122°F)
Adapter—p/n4240-403
Connector N (F) to Universal Adapter (M)
Impedance 50
Ώ nominal
Verification Data—Rho data included with mismatch,
Data Uncertainty
VSWR ± 2.4%
Rho ± 0.0121
ρ
VSWR 1
VSWR 1+
-------------------------=