User guide

124
Log: On / Off
Activates or deactivates the internal logging feature.
When logging is On, a thin line appears under the word On on the soft key text.
When logging is Off, a thin line appears under the word Off.
Logging captures and stores the set of data that is sent from the external power
sensor and displayed on the screen. While logging is On, each set of data is
stored as text on a separate line in the same computer file. The speed of the log-
ging depends on the type of sensor being used (e.g., a 5011 will record a data
every 300 ms).
Logging stops when the Log: On / Off soft key is pressed. The log file is named
using the date and time file naming convention PwrYYYYMMDDHHMMSS.csv
where YYYY is the year, the first MM is the month, DD is the day, HH is the hour,
the second MM is the minute, and SS is the second of the time when the file was
saved.
When the logging feature is On, there is a choice to log (store) every data set
received or one set from a group. Press the up or down arrow keys to change
the number of data sets to store:
1 of 1 - store every data set received
1 of 10 - store every tenth data set received
1 of 100 - store every one-hundredth data set received
1 of 1000 - store every one-thousandth data set received