SignalHawk Procedure
1. Navigate to the Utilities Menu.
2. Press “Exit to Windows” under the Utilities Menu.
3. Use the arrow keys to highlight the SignalHawk Diagnostics application on
the SignalHawk desktop.
4. Press Enter to start the operation.
5. Press the “Digital Board” key.
6. Press the “Update FPGA” key.
7. Navigate to the FPGA_xxx file (under “FlashFX Disk”) and select it.
8. Press Enter to start the operation.
9. Press Enter at each confirmation message until the Diagnostic Application
exits.
Note: This may take a moment.
10. Restart the Diagnostic Application.
11. Press the “Digital Board” key.
12. Then the “Update DSP” key.
13. Navigate to the DSP_xxx file (under “FlashFX Disk”) and select it.
14. Press Enter to start the operation.
15. Press Enter at each confirmation message until the Diagnostic Application
exits.
Note: This may take a moment.
16. Reboot the SignalHawk.