User guide

147
Power Sensors Supported
Sensor Description
5010B Frequency Range Element dependent, 2 MHz to 2.7 GHz
Power Range Element dependent, 125 mW to 1 kW full scale
Measurements performed: True Average Power, Peak
Forward Power (element dependent).
Calculations Performed: VSWR, Return Loss, Reflection
Coefficient
Accuracy True Average Power, ± 5% of reading
(15 °C to 35°C), ± 7% of reading (-10 °C to 50°C),
Peak Power, ±8% of full scale
Requires two Bird elements
5010T Frequency Range Element dependent, 2 MHz to 2.7 GHz
Power Range Element dependent, 125 mW to 1 kW full scale
Measurements performed: True Average Power, Peak
Forward Power (element dependent).
Calculations Performed: VSWR, Return Loss, Reflection
Coefficient
Accuracy True Average Power, ± 5% of reading
(15 °C to 35°C), ± 7% of reading (-10 °C to 50°C)
5011 Frequency Range 40 MHz - 4.0 GHz
Power Range -20.000 to +10.000 dBm
(10.0010 W to 10.000 mW)
Measurements performed: True average power.
Accuracy ± 5% of Reading. When operating below 100 MHz
and above 40 °C, add 1%.
5011-EF Frequency Range 40 MHz - 12 GHz
Power Range -20.000 to +10.000 dBm (10.0010 W to
10.000 mW)
Measurements performed: True average power.
Accuracy ± 5% of Reading. When operating below 100 MHz
and above 40 °C, add 1%
5012B Frequency Range 350 MHz - 4.0 GHz
Power Range 150 mW - 150 Watts Avg. 400 Watts
Peak
Measurements performed: Peak power, true average
power and Duty Cycle.
Calculations Performed: VSWR, Return Loss,
Reflection Coefficient, Crest Factor, Average Burst
Power and CCDF.
Duty Cycle (D) .001 to 1.0
Accuracy Average Power = ±4% of reading, + 0.05 Watts
Burst Average Power = ±6% of reading, +0.05/D Watts
Peak Envelope Power = ±7% of reading, +0.20 Watts
CCDF = ±0.2% typical