4.3 FILTER
- Set the attenuator switching response time. The cursor moves to the position indicated below when
the <FILTER> key is pressed. This parameter can be set in either [MANUAL] mode or [PRGM]
mode.
[MANUAL] mode example:
MANUAL 23.45dB
10μs
[PRGM] mode example:
PRGM 500KHz L:131072
P:6.5535s FREE 10μs
- Filter Response time can be entered in 1-3 steps in the range of 1 µs – 3 ms using the rotary knob.
Note
: Use the built-in filter to reduce switching transients. To set the filter response time appropriately in
relationship to the readout clock, a value of 1/2 to 1/8 of the clock period is recommended.
4.4 READ CLOCK
- Select the program memory read clock. The cursor moves to the position indicated below when the
<READ CLOCK> key is pressed. This parameter is valid in [PRGM] mode only.
PRGM 500KHz L:131072
P:6.5535s FREE 10μs
- [MANU], [EXT], [100Hz] - [500kHz] can be selected using the rotary knob.
① When [MANU] is selected, one clock cycle can be generated by pressing the <ENTER> key.
② When [EXT] is selected, the signal applied to terminal READ CLOCK IN is used as the clock.
③ When the range of [100Hz] - [500kHz] is selected, the internally generated clock is used.
4.5 PRGM LENGTH
- Set the program memory length. The cursor moves to the position indicated below when the
<PRGM LENGTH> key is pressed. This is valid in the [PRGM] mode only.