Specifications
11
PRGM 500KHz L:131072
P:6.5535s FREE 10μs
- Move the cursor using cursor keys, and set a value in the range of 8 - 131072 using the rotary knob.
It is provided, however, that the upper limit depends on the date length in the program memory.
- Press the < key to move towards the most significant digit and press the > key to mode towards the
least significant digit.
4.6 PAUSE TIME
- The pause time is defined as the interval between readout periods. The first pause time is the time
interval between the first and second readout period.
Figure 1: Readout period and Pause time
The screen below appears when the PAUSE TIME key is pressed. Highlight and select one of the
parameters [TIME], [CLOCK] or [OFF]. The current setup is displayed in the right upper part of the
screen. This parameter is valid in the [PRGM] mode only.
Note: During the Pause time, the last attenuation value in the program remains active.
① When [TIME] is selected, the main screen is displayed and the cursor moves to the position
indicated below.
PRGM 500KHz L:131072
P:6.5535s FREE 10μs
- Move the cursor using the cursor keys and set a value in the range of 0 s - 6.5535 s using the rotary
knob. The cursor moves towards the most significant digit when < is pressed and towards the least
significant digit when the > key is pressed. Values can be set with 100 µs resolution.
PAUSE: TIME 0.0000s
TIME CLOCK OFF