Model: XLN3640, XLN6024, XLN8018, XLN10014 High Power Programmable DC Power Supply USER MANUAL
Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases of operation, service, and repair of this instrument. Before applying power, follow the installation instructions and become familiar with the operating instructions for this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument.
DO NOT SUBSTITUTE PARTS OR MODIFY THE INSTRUMENT Do not install substitute parts or perform any unauthorized modifications to this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety features are maintained. WARNINGS AND CAUTIONS WARNING and CAUTION statements, such as the following examples, denote a hazard and appear throughout this manual. Follow all instructions contained in these statements.
Store/Move/Maintain Storage When this device is not in use, properly package it and store it in an environment suitable for storage (if present in a good preserving environment, the packaging process can be waived). Freight While moving this product, move it by using the original packaging to pack this product in advance. If the packaging material is lost, use an equivalent buffer material to replace it in packaging; and with external marks indicating “fragile & water-prevention”.
Table of Contents 1. Preface...................................................................................................... 1 1.1 Products Outline ................................................................................ 1 1.2 Features ............................................................................................. 2 1.3 Specifications .................................................................................... 4 2. Cautions Before Using ..............................
4.6 Control Voltage Output with Rotary knob ....................................... 34 4.7 Timer Function ................................................................................ 34 4.8 Series (cascade) / Parallel Mode Setting ......................................... 35 4.8.1 Parallel Connection Setting........................................................ 35 4.8.2 Series Mode Setting ................................................................... 37 4.8.
6.6 State Bit Definition .......................................................................... 73 6.7 LAN Communication (-GL verions) ............................................... 74 Using Web Server........................................................................................... 74 Main Page (Home) .................................................................................. 74 Using Telnet ............................................................................................
1. Preface 1.1 Products Outline B&K Precision models XLN3640/XLN6024/XLN8018/XLN10014 are programmable DC power supplies with single outputs that offer the maximum power output up to 1440 watts (0 -- 36 V/40 A or 0 -- 60 V/24 A or 0 -- 80 V/18 A or 0 -- 100 V/14.4 A). With a 16-bit D/A, A/D converter embedded, the power supplies come with the resolution of 1mV in voltage setting and 1mA in current setting.
1.2 Features 1) Output Voltage & Current Voltage output range: (XLN10014) Current output range: (XLN10014) Power output range: 0 -- 36V (XLN3640) / 0 -- 60V (XLN6024) 0 -- 80V (XLN8018) / 0 -- 100V 0 -- 40A (XLN3640) / 0 -- 24A (XLN6024) 0 -- 18A (XLN8018) / 0 -- 14.4A 0 -- 1440W 2) Rotary knob, numerical keys and functions keys The rotary knob can be used to rapidly change the output voltage setting and simulate the surge of the voltage output. It offers a good solution for testing triggering circuits.
UUT by accidentally entering the wrong settings. 6) Series & parallel connection mode The series-parallel connection mode of two or more units (maximum to 4 units) significantly increases the combined output power to a maximum of 5760 W. In parallel connection mode of four XLN3640 the maximum output is 36 V/160 A; and in series connection mode of four XLN10014, the maximum output is 400 V/14.4 A.
1.3 Specifications XLN3640/XLN6024 Specifications Output Rating Output Voltage XLN3640 0--36 V XLN6024 0--60 V Output Current 0--40 A 0--24 A Output Power Output Protection OVP Adjustment Range OVP Accuracy Line Regulation Voltage 1440 W 1440 W 2--38 V 200 mV 3--64 V 300 mV ≤ 4 mV ≤ 6 mV Current ≤ 4 mA ≤ 4 mA ≤ 8 mV ≤ 8 mA ≤ 8 mV ≤ 7 mA ≤ 5 mVrms/≤ 60 mVpp ≤ 6 mVrms/≤ 70 mVpp ≤ 90 mA ≤ 70 mA Programming 1 mV/1 mA 1.5 mV/1 mA Readback 1 mV/1 mA 1.
General Average Command Response Time <50 ms <50 ms 0.99 (Full load) 2V 0.
XLN8018/XLN10014 Specifications Output Rate XLN8018 XLN10014 Output Voltage 0~80 V 0~100 V Output Current 0~18 A 0~14.4 A Output Power 1440 W 1440 W OVP Adjustment Range 4~85 V 5~105 V OVP Accuracy 400 mV 500 mV Voltage ≤ 8 mV ≤ 10 mV Current ≤ 4 mA ≤ 4 mA Voltage ≤ 10 mV ≤ 12 mV Current ≤ 6.5 mA ≤ 6 mA Output Protection Line Regulation Load Regulation Ripple/Noise (20Hz-20MHz) Normal Mode Voltage ≤ 7 mVrms/≤ 80 ( ≥ 0.5 % of max.
General Average Command Response Time Power Factor Correction <50 ms <50 ms ≥ 0.99 (Full load) ≥ 0.
Features of models XLN3640/XLN6024/XLN8018/XLN10014: Graphical, easy to read LCD display Compact, high efficiency and power density 40 A output connector for quick connectivity Convenient numerical & function keys Store and recall 10 instrument settings Timer (1 sec -- 100 hours) Programmable (SCPI command only) List mode supports up to 10 sets of program and maximum 150 steps in total Auxiliary 5 V/1 A output Built-in precise voltage and current measurement OVP, OCP, OPP and key-lock
2. Cautions Before Using 2.1 Check and Confirm Accessories before Using After receiving this product, please verify the items received in accordance with the ones listed below: 1. The appearance of the products is without scratch or other damages. 2. Standard parts as shown in parts list of section 8. 2.2 Operation Instructions In order to avoid damaging the instrument due to improper operation, be sure to read this user manual.
2.4 Storage The storage temperature range of this product is within -10ºC - 70ºC and R.H. should be within 80% without moisture condensing. If not operating this product for a long time interval, pack it with original packaging or similar one and put it in a dry place without exposure to direct sunlight. 2.5 Power-line voltage Rated AC power source connected to this product is within 100 V-240 V (refer to the Product Specification for details).
2.7 Warm-up Time The XLN series is fully operable upon switching the power on. However, to reach the specified equipment accuracy, please allow the supply to warm up for at least 30 minutes. 2.8 Power-off procedure When the supply is not in use, be sure to turn the power switch on the panel to the OFF position to turn off the power.
3. Front Panel Operation 3.1 XLN3640/XLN6024/XLN8018/XLN10014 Panel 4 3.1.1 (1) 3 2 1 15 6 5 9 7 8 11 14 13 10 12 Front Panel Power switch: Please consult the “Cautions before use” section before turning on power switch. (2) Display: 192x32 Graphic LCD Module (3) Press (4) to set up the current limit. Voltage setting Press (5) : Current setting Dot/Local : to set up the output voltage. : This button is applied as a decimal point.
(6) ESC/CLR : Press this button to clean up numerical setting or jump to the previous screen. (7) Numerical keys - : They are used to directly input the voltage or current value or choose the setting option in Menu screen. (8) Down/Right/Store : This key is a multi-function key for the following three functions: (9) Down: In “Menu Setting” status, use this “Down” key to move cursor to the next item. Right: Under “Output” status, use this key to move cursor right.
= 15 . = 36 . . (11) Output . FF : Control the On/Off of the output power. (12) The rotary knob: Use this knob to adjust voltage or current (press cursor display first). This is adjustable when output is ON. (13) Enter first to let : This key is the confirmation key of current or voltage setting value; or press under output status to dynamically adjust voltage (at CV mode) or current (at CC mode). (14) Mem : Press this key to enter access the storage memory.
1 . SYSTEM SETTING 2 . OUTPUT SETTING 3 . PROTECTION 4. 5. 6. F T L T LL L T F 7 . TIMER CONTROL 8 . CALIBRATION 1. SYSTEM SETTING: Pressing key in the first page of Menu Setting will enter the following “SYSTEM SETTING” menu. T L= = L= 1 FF REMOTE CONTROL: Choose the remote interface (USB/GPIB/ETHERNET) *GPIB and ETHERNET available only with on models with “-GL” suffix *USB control requires installing USB drivers first. Download USB driver from www.bkprecision.
Data bit :8 Parity check : none Stop bit :1 *When entering the Remote mode, screen will present RMT indicator as shown in the following picture. = 15 . = 36 . . . FF GPIB ADDRESS: Set up GPIB ADDRESS (1-31) EXTERN CONTROL: Set up the external control to voltage control (VOLT 0-10 V or 0-5 V), resistance control (RES 0-5K) or off (OFF). F = = 255 . 255 . 255 . 255 = FF L IP CONFIG: STATIC : User can input IP address DHCP : IP address is assigned by the server.
L L LL F = = L = L BEEP: Turns the Buzzer ON/OFF LCD BACKLIT: Set the backlight of the LCD to Always ON or OFF after 1/5/10/30 minutes RECALL DEFAULT: Restores settings 5 = , = . = = the . manufacturer FF FF , = default FF Ext 5V OUTPUT: Turns the extra 5V power output (on the rear panel) ON/OFF. POWER ON STATE: Users can set the output state of the supply when powered on. When OFF is selected, the XLN series will do nothing after power on.
L F LL 2.5 = = L = 1 L VOLT LIMIT: Upper limit of the output voltage setting CURR LIMIT: Upper limit of the output current setting VOLT SLEW RATE: Voltage ascending/descending slope (XLN3640: 0.01 - 2.4V/ms) (XLN 6024:0.01 - 3V/ms) (XLN 8018:0.01 - 3.2V/ms (XLN 10014:0.01 - 3.3V/ms) CURR SLEW RATE: Current ascending/descending slope (XLN 3640:0.01 - 2.5A/ms) (XLN 6024:0.01 - 1.2A/ms) (XLN 8018:0.01 - 0.72A/ms (XLN 10014:0.01 - 0.
OVP:turns on/off the overvoltage protection OCP:turns on/off the overcurrent protection OPP:turns on/off the overpower protection = = SET: set up the overvoltage protecting point. SET: set up the overcurrent protecting point. SET: set up the overpower protecting point. FF FF Enable/disable the protection of the change from CV to CC mode Enable/disable the protection of the change from CC to CV mode CV TO CC: CC TO CV: 4.
L 364 L F : 1 . 13 L 6. SPECIAL TEST FUNCTION Press in the second screen of “Menu Setting” to enter SPECIAL TEST FUNCTION menu. 1. 2. 6.1 CURRENT COUNTER TEST: Press CURRENT COUNTER TEST screen. = 1 . : = 1. : . = to enter the . FF Refer to “Current Counting” section for the detailed setting procedure. to enter the PROGRAM MODE 6.2 PROGRAM MODE: Press menu.
the program. 7. TIMER CONTROL in the third page of Menu Setting to enter TIMER Press CONTROL screen. = FF = r in ec TIMER: Turn on/off TIMER function. TIME: Set up OUTPUT ON time (Max:999Hr 59Min 59Sec) 8. CALIBRATION in the third page of “Menu Setting” to enter Press CALIBRATION menu. Users must enter the password to access calibration mode. : L 8.1 Equipment Requirements 1. 5 ½ Digital Volt meter. 2. Shunt for current calibration (100 A/ 10 mΩ) 8.
A. Connect power supply output terminal to DVM (as shown in Figure 1 below). Turn on the supply. Once the unit enters the main page, press MENU and select “8. Calibration” and key in password “13579” to enter the following calibration menu screen: Figure 1 1. 2. 3. L L L L B. Press “1” to access Voltage Calibration Procedure. C.
L L L L L L Lo = 1 . 864 L= 12 . 554 = 2 5 11 4 L L i = 35 . 78 D. According to voltage value displayed on DVM, fill in the values for their corresponding functions and press ENTER. If any DVM read-back value at each voltage function does not fit with the following table below, please inspect the hardware. XLN3640 Function Setting Range of Read-Back Value Value Lo 1.8V 1.5 - 2.0 V MIDL 12V 10 - 13 V MIDH 24V 21 - 25 V Hi 32.
Value Lo 4V 3.6 - 4.4 V MIDL 26V 23.5 - 28.5 V MIDH 53V 48 - 58 V Hi 76V 69 - 83 V Setting Range of Read-Back Value XLN10014 Function Value Lo 5V 4.5 - 5.5 V MIDL 33V 29.7 - 36.3 V MIDH 66V 59 - 72 V Hi 95V 85 - 104 V E. When pressing ENTER at Hi function, and the calibration data values will be stored in FLASH and unit will return to calibration main screen. OVP CALIBRATION A. Press 2 in the calibration main page to enter OVP calibration page. L L Lo = i = a B.
L L Lo = i = al ib a in . . . C. Jump to Hi function when completed. L L Lo = i = a D. Press ENTER to access OVP calibration procedure at High function. E. L L Lo = i = al ib a in . . . F. After completing the above steps, calibration value will be stored in FLASH and unit will return to the calibration main screen. G. If calibration is not completed in 10 seconds after starting, please inspect OVP circuit. CURRENT CALIBRATION A.
B. First, input parameter (resistance of current shunt, mΩ) of current measurement fixture. According to voltage value shown on DVM, fill in that value that corresponds to the function and press ENTER key.
F L L = 1 . Lo = 1. L= 13 . . L L Li L= 2 6 . = 38 . C. Please inspect the hardware if the current value is inconsistent with the value in the tolerance error table below after the calibration values of each function is changed to actual current value. XLN3640 Function Setting Range of Transformation Value Lo 0.1A 0 - 0.5 A MIDL 13A 11 - 14 A MIDH 26A 23 - 27 A 38A 34 - 39 A Hi XLN6024 Function Setting Range of Transformation Value Lo 0.06A 0 - 0.5 A MIDL 7.8A 6.
MIDH Hi 12A 10.8 - 13.2 A 17A 15.3 - 18.7 A XLN10014 Function Lo Setting Range of Transformation Value 0.03625A 0 - 0.1 A MIDL 4.7A 4.2 - 5.2 A MIDH 9.5A 8.55 - 10.45 A 13.7A 12.33 - 15.07 A Hi D. The calibration value will be stored in FLASH after pressing ENTER at Hi function. 9. Series Connection Control Setting (CHAIN) Press page.
3.1.2 Rear Panel 16 17 18 19 26 20 21 22 25 24 23 Cooling fans: The supply automatically adjusts fan’s rpm according to the load condition. (16) Cooling fan: The rear cooling fan speed is temperature control. (17) Power output terminal: Please pay attention to the correct polarities when making connection.
When Local sense is selected, the wire connection can be setup as follows (shown in the figure above):positive sense (+S) is connected to the positive lead (+) and negative sense (-S) is connected to the negative lead (-), whereas the positive lead (+) of the DC output is connected to the positive end (+) of the device and the negative lead (-) of the DC output is connected to the negative end (-) of the device under test.
(23) 5V/1A Output: XLN series offers an extra output with a constant output voltage of 5 V and the maximum output current of 1 A. This extra power supply can be switched on or off under the “System Setting” menu. (24) USB : USB interface connector. (25) EXT CTL: Models XLN3640/XLN6024/XLN8018/XLN10014 offer the capability of setting the output voltage/current by adjusting an external input voltage/resistance.
WARNING: This PRODUCT is designed meeting safety code and has passed the related qualification test. In case no output cable is connected, close the protection cover and fasten the screws in so as to protect user from electric shock or other hazards.
4. Operation Instructions 4.1 Voltage Setting Press and set the output voltage by pressing the numerical keys directly, and then press to confirm the setting. = 4.2 = 15 . . . FF Current Setting Press and set the output current by pressing the numerical keys directly and then press to confirm the setting. = 36 . = . 4.3 . FF Over-voltage Protection OVP to enter the Configuration menu and press to enter the Press PROTECTION setting menu.
to confirm it. Now the cursor will move to the value setting for the OCP on the right hand side. Enter the OCP value by pressing the numerical keys. = = = 4.5 FF S S S 38 . = 42 . = = 1440 . Voltage Output After the voltage, current and protection settings are entered, press to output voltage. User will be able to recognize the setting values and the actual output values from the LCD. = 36 . = 15 . . . 4.
countdown of the timer. Once it reaches down to zero, the supply will turn off the output automatically. . 4.8 = : : 55 . ec Series (cascade) / Parallel Mode Setting The XLN3640/XLN6024/XLN8018/XLN10014 uses the series/parallel mode (4 supplies maximum) to increase the output power capability. By connecting 4 instruments in parallel, the combined unit can offer 36 V/160 A power output. By connecting 4 XLN10014 units in series, the combined supply can output 400 V/14.4 A.
that are connected to the Master. Therefore, in order to correctly configure the correctly, user must set up the Slaves before the Master is set. To set a XLN3640/XLN6024/XLN8018/XLN10014 to slave mode, press in the main screen to enter the series/parallel , , and setting option. And then select the parallel mode by turning the knob (SCPI to confirm. It command is “PS:MODE PARALLEL”) and then press will continue to the next line for the MASTER/SLAVE selection.
After receiving the control command from the Master, all Slaves will be locked are locked so on to the SLAVE screen. At this point all keys except that slaves cannot be operated by front keys except controlled by the Master as shown below: : LL L : Once the Master and Slave settings are done, users can then operate the Master only to set up the combined output voltage and current of the complete system.
After wiring is complete, configure one supply as the Master and the other three as the Slave A, B, and C. After the power supply is configured as Master, it will start searching for all Slaves that are connected to the Master. Therefore, in order to setup correctly, user must set up the Slaves before the Master is set. To set a XLN3640/XLN6024/XLN8018/XLN10014 to the slave mode, press in the main screen to enter the series/parallel , , and setting option.
start searching for all Slaves, as shown below. L L = = F L ... If wiring is correct, this screen will appear: L F L L : = = After receiving the control command from the Master, all Slaves will be locked are disabled so that slaves on the SLAVE screen and all keys except cannot be operated by pressing front keys. This ensures the Master supply be in control. : : Once the Master and Slave settings are done, users can then operate the Master only to set up the output voltage and current.
display the following message in the screen for searching Slaves: L F L L : = = LL L If more than one Master is being set in the series/parallel operation mode, the following message will be displayed. L L L = = LL L , L If one of the power supplies is set to different series/ parallel operation mode, the following message will appear on-screen:.
: 4.9 LL L : External Tuning Setting Models XLN3640/XLN6024/XLN8018/XLN10014 provides the capability to control the output voltages by an external voltage/resistance.
10mV): V = 36.00 V I = 40.00 0.00 V 4.10 A 0.00 A OFF Timer of Current Flow This function offers testing of the cutoff time of a breaker or a fuse. First, connect the test object to output terminal as shown below. , and then Press , , current flow counter screen. = 1 . : = 1. : in the main screen to enter the . = . FF After setting up the output voltage/current and pressing to turn on the output, the system will start counting down the timer until the breaker or fuse is open.
message as shown below. F : L! ! : ! . B. Current setting too large The output current is set too big that a breaker is open or a fuse is burnt to open before output current is reaching the setting value. The screen will show an error message as shown below. F : L! ! : . L ! C. Voltage setting too small The voltage is set too small, therefore the output current is unable to reach the setting value. The screen will show an error message as shown below. F 4.11 : L! ! : .
minimum) to stay on the step. Please refer to “SCPI command subsystem” section for detail SCPI commands. Below are some examples of commands used to setup a custom program in list mode.
PROG:STEP 4 PROG:STEP:CURR 1 PROG:STEP:VOLT 20 PROG:STEP:ONT 0.1 PROG:STEP 5 PROG:STEP:CURR 1 PROG:STEP:VOLT 15 PROG:STEP:ONT 0.1 PROG:STEP 6 PROG:STEP:CURR 1 PROG:STEP:VOLT 10 PROG:STEP:ONT 0.1 PROG:STEP 7 PROG:STEP:CURR 1 PROG:STEP:VOLT 5 PROG:STEP:ONT 0.1 PROG:STEP 8 PROG:STEP:CURR 1 PROG:STEP:VOLT 0 PROG:STEP:ONT 0.
Example 2: To output the wave form shown above, the following example program can be used. PROG 2 Choose program number PROG:CLE Clear program 2 data PROG:REP 0 No repeat after running this program PROG:TOTA 8 Set program 2 to have 8 steps in total PROG:STEP 1 Settings for step 1 PROG:STEP:CURR 2 Set output current to 2 amperes PROG:STEP:VOLT 20 Set output voltage to 20 volts PROG:STEP:ONT 0.5 Set output ON time to 0.5 sec PROG:STEP 2 Choose step 2 PROG:STEP:CURR 2 PROG:STEP:VOLT 15 PROG:STEP:ONT 0.
PROG:STEP:ONT 0.5 PROG:STEP 6 PROG:STEP:CURR 2 PROG:STEP:VOLT 5 PROG:STEP:ONT 0.5 PROG:STEP 7 PROG:STEP:CURR 2 PROG:STEP:VOLT 20 PROG:STEP:ONT 0.5 PROG:STEP 8 PROG:STEP:CURR 2 PROG:STEP:VOLT 0 PROG:STEP:ONT 0.
Example 3: If it needs to execute Program 2 right after Program 1 is executed then program 1 shall be modified to have NEXT 2 command. The following steps can be taken for the modification and execute the programs. PROG 1 Select program 1 PROG:NEXT 2 Change the NEXT command from NEXT 0 to NEXT 2 PROG:SAV After edit is complete use Save command to store changes in the hardware PROG 1 To run the program, select the program number first and then use RUN ON command to execute it.
4.12 Multi-unit Connection mode (RS485) XLN3640 / XLN6024 / XLN8018 / XLN10014 can use RS485 to provide multi-units series connection function for up to 31 units (If more than 10 units, please add a 120Ω resistor terminator in the last unit as shown in the below figure. Turn on the system after series connection is completed. Press on the main page and set CHAIN ON/OFF to ON (Series/Parallel connection will be cancelled) and set each unit with a different Address (1 - 31).
SYSTEM CONTROL COMMAND: Command Description ================================================== CADR followed by address, which can be 1 to 31 and is used to access the power supply CCLS Clear status CRST Reset command.
COC? COCP COCP? COP COP? COPP COPP? CMODE? Returns the OCP setting level Sets the OCP on/off Returns the OCP on/off Sets the OPP level Returns the OPP setting level Sets the OPP on/off Returns the OPP on/off Returns the power supply operation mode ================================================== SYNCHRONOUS CONTROL COMMAND: Command Description ================================================== GRST Reset command.
OK OK CADR 7 CPV 20 Q3. How to set up Output for Address 7 on the system? OK OK CADR 3 COUT 1 Q4. How to read back Voltage value for Address 1 on the system? OK 10.001 CADR 1 CMV? Q5. How to set up Current for all the systems? No response GPC 5 Q6.
5. Protection and Error Messages 5.1 Over-voltage Protection (OVP) When the OVP is activated and voltage measured exceeds the setting point of protected voltage, the system will enter the “Over Voltage Protect” mode that will shut off the output and show OVP on the display. Press to reset the protection mode and deactivate the buzzer. = 36 . = 40 . . 5.2 .
5.4 Constant Voltage Protection (CV TO CC) When this function is activated, the power supply will stay in CV mode. If load changes force the system to transition from CV to CC (constant current) mode, the system will enter the “CV TO CC Protect” state that will shut off the output and display the CVC message on the screen. Press to reset the protection and deactivate the buzzer. V = 36.00 V I = 40.00 0.000 V 0.000 A 5.
5.7 Low Voltage Protection (ACD) When the machine has detected abnormally low AC power input, system will enter the “AC Detect Low Protect” mode that will shut off output and display the error message shown in the following figure. Press to reset the protection and deactivate buzzer. L 5.8 L L !!!! . . Error Input Message When users enter a voltage or current setting that is beyond the acceptable range, system will display “RANGE ERROR” in response and show users the correct input range.
6. Remote Interface communication protocol The communication protocol includes standard SCPI commands and a few proprietary commands which follow the SCPI convention. 6.1 Prefaces The SCPI interface enables users to operate the model XLN3640/XLN6024/XLN8018/XLN10014 supply through a computer or a terminal equipped with IEEE-488.2 GPIB or USB interface. Additionally, it allows remote control and monitoring. SCPI IEEE-488.
characters at the end of each command when programming. 6.3 The Error/Event List SCPI interface can offer an error/event list that contains up to 10 errors/events. Users can read the errors/events through the “error?” command in a first-in first-out manner. Once an error/event is read, the read process will clear it from the memory. To clear all errors/events from the memory, the “*CLS” command is used.
6.
Examples: Q 1:How to set GPIB address? ADDR 10 ==>address is 10 Q 2:How to read back GPIB address? ADDR? ==>return GPIB address ADDRESS? ==>return GPIB address Q 3:How to set up buzzer? BEEP 1 ==>trigger beep to on BEEP off ==>trigger beep to off Q 4:How to clear the protecting state? CLR ==>clear protect status Q 5:How to read back to error information? ERR? ==>return error code Q 6:How to set up voltage? VSET 10 ==>set voltage to 10V VOLT 3.3V ==>set voltage to 3.
Q 11:How to read the current outputting value? IOUT? ==>return current output Q 12:How to lock buttons and the rotary knob? LOCK 1 ==>lock the keypad and knob LOCK ON ==>lock the keypad and knob Q 13:How to read the product model number? MODEL? ==>return machine model name Q 14:How to set up OVP function? OVP 1 ==>enable OVP protect OVP OFF ==>disable OVP protect Q 15:How to set up OVP voltage value? OVSET 38 ==>set OVP level to 38 V Q 16:How to set up OCP function? OCP 1 ==>enable OCP protect OCP OFF ==>di
6.5 SCPI Conformity Information The XLN series is entirely compatible with all IEEE-488.2 and SCPI commands (1995). Commands are included in SCPI commands (in Vol. 2, 1995). 6.5.1 Common SCPI commands Command Explanation *CLS Clear status (include error code) *IDN? Response:, , , *RCL Recalls settings from memory. Memory numbers from 0 to 9 are valid. *RST Resets the power supply to its power on state. *SAV 1.
*CLS 6.5.
from 1 - 10 :CLEar :ALL clear all program parameters :NEXT ser or return next program number (1 10 , 0 for end) :REPeat set or return repeat times (max. 50000) :RUN set or query program on/off state :SAV save program parameters :STEP < Boolean | ? > set or return step number :CURRent set or return step n current setting :ONTimet set or return step n output time (0.
:MODE set or return parallel/series mode :TYPE set or return master/slave setting [SOURce] Source Subsystem :CURRent set or return current level:0-40/0-24A :PROTection set or return over-current state :LEVel set or return over-current level :VOLTage set or return voltage level:0-36/0-60V :PROTection set or return over-voltage state :LEVel set or return over-v
:LOCK set or return LCD backlight state :POWer :CURRent set or return user define current level :STATe set or return user define output state :TYPE set or return power up mode :VOLTage set or return user define voltage level :RECall :DEFault recall factory default setting :REMote set or return remote interface :SERies? TIMER Return seri
Examples: Q 28:How to cancel all actions? ABOR ABORT Q 29:How to fetch current value? FETC:CURR? FETCH:CURRENT? Q 30:How to fetch voltage value? FETC:VOLT? FETCH:VOLTAGE? Q 31:How to measure current? MEAS:CURR? MEASURE:CURRENT? Q 32:How to measure voltage? MEAS:VOLT? MEASURE:VOLTAGE? Q 33:How to set up and read back the specific memory set? MEM 1 MEMORY 3 MEM? MEMORY? Q 34:How to set up and read back the voltage stored in specific memory set? MEM:VSET 10 MEM:VSET? MEMORY:VSET 20 MEMORY:VSET? Q 35:How to set
MEM:ISET? MEMORY:ISET 25 MEMORY:ISET? Q 36:How to save data to memory set? MEM:SAVE MEMORY:SAVE Q 37:How to set up and cancel output? OUT ON OUTPUT 0 Q 38:How to set up and read back the voltage limit? OUT:LIM:VOLT 30 OUT:LIM:VOLT? OUTPUT:LIMIT:VOLTAGE 35 OUTPUT:LIMIT:VOLTAGE? Q 39:How to set up and read back the current limit? OUT:LIM:CURR 30 OUT:LIM:CURR? OUTPUT:LIMIT:CURRENT 35 OUTPUT:LIMIT:CURRENT? Q 40:How to set up and read back the voltage SLEW RATE? OUT:SR:VOLT 2.4 OUT:SR:VOLT? OUTPUT:SR:VOLTAGE 0.
Q 43:How to read back the protection state? PROT? PROTECTION? Q 44:How to set up and read back the CC to CV protection state? PROT:CCCV ON PROT:CCCV? PROTECTION:CCCV 0 PROTECTION:CCCV? Q 45:How to set up and read back the CV to CC protection state? PROT:CVCC ON PROT:CVCC? PROTECTION:CVCC 0 PROTECTION:CVCC? Q 46:How to clear the state of protection? PROT:CLE PROTECTION:CLEAR OUT:PROT:CLE OUTPUT:PROTECTION:CLEAR Q 47:How to set up and read back the overcurrent protection state? PROT:OCP ON PROT:OCP? PROTECTIO
SOURCE:CURRENT:PROTECTION:LEVEL? Q 49:How to set up and read back the overpower protection state? PROT:OPP ON PROT:OPP? PROTECTION:OPP 0 PROTECTION:OPP? Q 50:How to set up and read back the overpower protection point? PROT:OPP:LEV 30 PROT:OPP:LEV? PROTECTION:OPP:LEVEL 40 PROTECTION:OPP:LEVEL? Q 51:How to set up and read back the overvoltage protection state? PROT:OVP ON PROT:OVP? PROTECTION:OVP 0 PROTECTION:OVP? SOUR:VOLT:PROT ON SOUR:VOLT:PROT? SOURCE:VOLTAGE:PROTECTION 0 SOURCE:VOLTAGE:PROTECTION ? Q 52:H
SYSTEM:E5V 0 Q 55:How to read back error code? SYS:ERR? SYSTEM:ERROR? Q 56:How to set up and read back the external adjustment state? SYS:EXT VOLT SYS:EXT? SYSTEM:EXTERNAL RESISTANCE SYSTEM:EXTERNAL? Q 57:How to set up and read back the GPIB address? SYS:GPIB:ADDR 5 SYS:GPIB:ADDR? SYSTEM:GPIB:ADDRESS 6 SYSTEM:GPIB:ADDRESS? Q 58:How to set up and read back the IP address? SYS:IP:ADDR 192.168.0.208 SYS:IP:ADDR? SYSTEM:IP:ADDRESS 192.168.10.
SYS:POW:TYPE? SYS:POWER:TYPE USER SYS:POWER:TYPE? SYSTEM:GPIB:ADDRESS? Q 63:How to set up and read back the voltage under user-defined booting mode? SYS:POW:VOLT 10 SYS:POW:VOLT? SYSTEM:POWER:VOLTAGE 20 SYSTEM:POWER:VOLTAGE? Q 64:How to set up and read back the current under user-defined booting mode? SYS:POW:CURR 10 SYS:POW:CURR? SYSTEM:POWER:CURRENT 20 SYSTEM:POWER:CURRENT? Q 65:How to set up and read back the output state under user-defined booting mode? SYS:POW:STAT ON SYS:POW:STAT? SYSTEM:POWER:STATE 0
Q 69:How to set up and read back the output voltage? SOUR:VOLT 30 SOUR:VOLT? SOURCE:VOLTAGE 35 SOURCE:VOLTAGE? Q 70:How to set up and read back the output current? SOUR:CURR 40 SOUR:CURR? SOURCE:CURRENT 35 SOURCE:CURRENT? Q 71:How to set up and read back parallel/series mode? PS:MODE PARALLEL PS:MODE? PS:MODE 2 PS:MODE OFF Q 72:How to set up and read back master/slave in parallel/series mode? PS:TYPE MASTER PS:MODE? PS:MODE SLAVEB PS:MODE 3 Q 73:How to set up Program function? Refer to “Programmable capabil
6.6 State Bit Definition When “STATE?” command is used, the system will return three (3) bytes in the order as shown below.
6.7 LAN Communication (-GL verions) XLN series power supplies provide three LAN control modes, including Web server, Telnet and Sockets. First, enter the first Item “System Setting” on the menu to set the interface selection (Remote Control) to Ethernet. Then, select IP setting procedure as Static or use DHCP (Dynamic Host Configuration Protocol). If using static IP, an IP address is required to complete setup.
Setting Page (Configuration) This page allows you to setup protection settings and reset/setup password for the system.
Status Page (Status) Display the status of the system. If any error codes are shown, please refer to “Error/Event list” section for details. Web Control This page allows user to set up voltage/current and output. Additionally, the display will allow monitoring of the output voltage/current value (JAVA support is required for display).
Using Telnet The power supply can be controlled via Telnet over the Ethernet interface. Simply go into command prompt under MS-DOS and enter: Telnet 5024 ( is IP address of the unit; 5024 is the Telnet port). A welcome message will be shown as below: Input SCPI command to communicate with the system, such as below: Using Sockets All XLN series use port 5025 to provide the standard function for SCPI socket.
7. Assemble Accessories 7.1 Assemble Rack Mount Brackets The XLN3640/XLN6024/XLN8018/XLN10014 is designed to fit in a space of one rack unit (1U) and can be mounted in a standard 19-inch rack panel or cabinet. Rack mount brackets must be assembled before mounting the unit in a rack. Refer to the following figure to assemble the rack mount brackets.
7.2 Assembly of Output Protective Cover (XLN6024/XLN8018/XLN10014) Safety regulation requests protective cover when the output voltage is higher than 45 VDC. Therefore, there are protective covers for the output connectors for the XLN6024/XLN8018/XLN10014. Refer to the following figure for the assembly/disassembly of the output protective cover.
7.3 Assembly of Remote Sense Protective Cover (XLN6024/XLN8018/XLN10014) Safety regulation requests protective cover when the voltage is higher than 45 VDC. Therefore, there are protective covers for the remote sense connectors for the XLN6024/XLN8018/XLN10014. Refer to the following figure for the assembly/disassembly of the remote sense protective cover.
8. Accessories Product Name: XLN3640/XLN6024/XLN8018/XLN10014 Parts List: Item Name & Specification 1. Remarks Terminal Block for Rapid Plug Connector 2. XLNRC Output Rapid Connector Includes two solder plugs 3. Rack mount Bracket kit 4. Handle bar kit 5. Operation Manual 6. XLNPC Power Cord 7.
9. Service Information Warranty Service: Please go the support and service section on our website www.bkprecision.com to obtain a RMA #. Return the product in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Non-Warranty Service: Please go the support and service section on our website www.bkprecision.com to obtain a RMA #.
10. Limited Two-year Warranty B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of two years from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt. To obtain warranty coverage in the U.S.A.
22820 Savi Ranch Parkway Yorba Linda, CA 92887 www.bkprecision.com © 2009 B&K Precision Corp. Printed in Taiwan v1.