User manual

21
the program.
7. TIMER CONTROL
Press
in the third page of Menu Setting to enter TIMER
CONTROL screen.
=
=
FF
r
ni ec
TIMER:
Turn on/off TIMER function.
TIME:
Set up OUTPUT ON time (Max:999Hr 59Min
59Sec)
8. CALIBRATION
Press
in the third page of Menu Setting to enter
CALIBRATION menu. Users must enter the password to access
calibration mode.
:
L
8.1 Equipment Requirements
1. 5 ½ Digital Volt meter.
2. Shunt for current calibration (100 A/ 10 mΩ)
8.2 Calibration Procedure
VOLTAGE CALIBRATION