A. Connect power supply output terminal to DVM (as shown in
Figure 1 below). Turn on the supply. Once the unit enters the
main page, press MENU and select “8. Calibration” and key in
password “13579” to enter the following calibration menu screen:
Figure 1
L1
.
3
2
.
.
L
L
L
B. Press “1” to access Voltage Calibration Procedure.
C.