Instruction manual

BIT DEFINITIONS-QUESTIONABLE DATA REGISTER
BIT DESCRIPTION DECIMAL
VALUE DEFINITION
0 OVP trip 1 Channel 1 Over Voltage Trip
2 OT trip 4 Channel 1 Over Temperature Trip
3 OVP trip 8 Channel 2 Over Voltage Trip
5 OT trip 32 Channel 2 Over Temperature Trip
6 Not Used 64 Always Set to 0
7 Not Used 128 Always Set to 0
8 OVP trip 256 Channel 3 Over Voltage Trip
10 OT trip 1024 Channel 3 Over Temperature Trip
11 OVP trip 2048 Channel 4 Over Voltage Trip
13 OT trip 8192 Channel 4 Over Temperature Trip
14 Not Used 16384 Always Set to 0
15 Not Used 32768 Always Set to 0
The questionable data event register is cleared when:
* You execute a *CLS (clear status) command.
* You query the event register using STATus: QUEStionable: EVENt?
The questionable data enable register is cleared when:
* You turn on the power to the instrument.
* You execute the STATus: PRESet command
* You execute the STATus: QUEStionable: ENABle 0 command.
6.6 COMMAND SUMMARY
This section summarizes the SCPI commands available for this Programmable
Instrument. Refer to the later sections in this chapter for more complete details on each
command.
Throughout this manual, the following conventions are used for SCPI command
syntax. Square brackets ([ ]) indicate optional keywords or parameters. Braces ({})
enclose parameters within a command string. Triangle brackets (< >) indicate that
you must substitute a value for the enclosed parameter.
44