Technical data

BKI 03 ATEX 027X spa3804a0600p_06
27 / 44
P34: - cba Logging mode
a Operating mode Parameters to be programmed
0 No logging
1 Linear logging P35 – interval (minute)
2 Event-controlled logging when the primary value changes P35 – absolute value of variation
3 Event-controlled logging when the primary value changes P35 – variation in %
4 Event-controlled logging when the primary value gets
out of range
P35, P36 – absolute values of range
limits
b Logging of errors and warnings (a>0)
0 No logging
1 Logging of all errors and warnings
2 Logging of errors only
3 Logging of NoEcho only
c Logging of changes in status (a>0)
0 No logging
1 Logging of change in status
FACTORY SETTING: 000 (no logging)
Errors that may generate an entry if P34/b<>0: NOECHO, ERR12, ERR13, ERR14, ERR15, ERR16, SUB0, ERR4, ERR5, PT ERR (temperature
measurement error).