P35-36: Log value 1 and log value 2
P34a Operating mode Function of P35 and P36
0 No logging
1 Linear logging P35 = 0 One entry after every measurement cycle.
P35 <> 0 Logging interval (minutes)
P36 Value is indifferent
2 Event-controlled logging when the primary
value changes
P35 Absolute value of the change in the primary measured value (according to
P01a)
P35 is defined as the dimension of the primary measured unit.
Log entry occurs if the value of the primary measured value deviates from the
measured value of the previous cycle in any direction by the amount
specified in P35.
P36 Value is indifferent
3 Event-controlled logging when the primary
value changes
P35 Relative (%) value of primary measured value (according to P01a) variation.
P35 is a relative quantity so its dimension is percent.
Log entry occurs if the value of the primary measured value deviates from the
measured value of the previous cycle in any direction by the amount
specified in P35.
P36 Value is indifferent
4 Event-controlled logging when the primary
value gets out of range in any direction
P35, P36 Absolute values of the limits of the range of the primary measured values
(according to P01a). P35 and P36 have the same dimension as the
primary measured value.
Log entry occurs if the value of the primary measured value steps
over the limit given in P35 and P36, in any direction.
To monitor only one limit value, set it in P35 and
let P36 be 0.
FACTORY SETTING: P35 = 0, P36 = 0