Technical data

BKI 03 ATEX 027X spa3804a0600p_06
41 / 44
7. ERROR CODES
Error
Code
Error description Causes and solutions
1 Memory error Contact local agent
No Echo Echo loss See Action 5 and 6
3 Hardware error Contact local agent
4 Display overflow Check settings
5
Sensor error or improper installation/mounting, level in
the dead band
Verify sensor for correct operation and check for correct mounting
according to the User’s Manual
6 The measurement is at the reliability threshold Better location should be found.
7
No signal received within the measuring range specified
in P04 and P05
Check programming, also look for installation mistake
12
Linearisation table error: both L(1) and L(2) are zero
(no valid data-pairs)
See ”Linearisation” Section
13
Linearisation table error: same L(i) data is given twice in
the table
See ”Linearisation” Section
14
Linearisation table error: the r(i) values are not
monotone increasing
See ”Linearisation” Section”
15
Linearisation table error: measured Level is higher than
the last Volume or Flow data-pair
See ”Linearisation” Section”
16 The check sum of the program is wrong Contact local agent
17 Parameter consistency failure Check programming
18 Hardware failure Contact local agent