User's Manual

12
Figure 9. Storage compartment door removal
5. Loosen the hinge on the other side and pull the hinge and lid free.
Save the lid and hinge for future use.
6. Disconnect all cables and remove them from the storage compartment.
7. On the back of the Merlin PSA, locate the gray Merlin PSA connector and connect it to the
gray RF Telemetry port in the storage compartment.
8. Locate the blue telemetry wand connector on the back of the Merlin PSA and connect it to
the blue telemetry wand port in the storage compartment.
9. Locate the black ECG cable on the back of the Merlin PSA and connect it to the black ECG
port in the storage compartment.
10. Orient the Merlin PSA over the storage compartment so its ports and LEDs are facing up and
the ports are facing away from the Merlin PCS screen. Gently lower the Merlin PSA into the
storage compartment.
11. Facing the back of the programmer, tilt the right side of the Merlin PSA into the
compartment so that the hinge hooks grip inside the storage compartment.
12. Push down on the Merlin PSA to snap it into place. It should fit securely and flush with the
surface of the Merlin PCS.