If the Merlin PSA Model EX3100 is already placed in the storage compartment, locate the
Merlin Antenna USB adapter (Model EX3190) provided with the Merlin PSA. Connect the USB
connector on the adapter to one of the USB ports on the left side of the Merlin PCS. Then,
connect the Merlin Antenna to the remaining connector on the adapter.
NOTE: Connect the Merlin Antenna only to the USB ports on the left side of the programmer.
Do not connect the Merlin Antenna to the USB port in the flash drive compartment.
Figure 12. Connection between the Merlin Antenna and the Merlin PCS using the Model EX3190
cable adapter
1. Merlin Antenna
2. Merlin PCS
3. Model EX3190 cable adapter
1. Place the Merlin Antenna on a flat surface approximately 1 – 2 m (2 – 6 ft) from the
implanted device. The front panel should face the device.
2. Follow steps 1 through 7 in the Merlin PCS Setup (page 7).
3. When the Merlin PCS is turned on, the telemetry strength indicator LEDs and the status
LEDs on the Merlin Antenna will light, indicating the unit is powered and operational.