disrupt the operation of nearby equipment. It may be necessary to mitigate this effect
by reorienting or relocating the receiving device or shielding the location.
Table 3. Guidance and manufacturer's declaration — electromagnetic emissions
Test
Compliance
Electromagnetic Environment Guidance
RF Emission
CISPR 11
Group 1
The Merlin™ PCS and the Merlin™ Antenna use RF
energy only for their internal function. Therefore, their
RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class A
The Merlin PCS and the Merlin Antenna are suitable for
use in all establishments other than domestic and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
The Merlin PCS and the Merlin Antenna are suitable for
use in all establishments other than domestic and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Voltage
fluctuations/flicker
emissions IEC
61000-3-3
Complies
Table 4. Guidance and manufacturer's declaration — electromagnetic immunity
Test
IEC 60601 Test Level (Actual
Level)
10
Electromagnetic Environment
Guidance
Electrostatic Discharge
(ESD) IEC 61000-4-2
±8 kV contact
(±8 kV contact)
±15 kV air
(±15 kV air)
Floors should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
11
±1 kV for input/output lines
(±1 kV for input/output lines)
±2 kV for power supply lines
(±2 kV for power supply lines)
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line[s] to line[s]
(±1 kV line[s] to line[s])
±2 kV line[s] to earth
(±2 kV line[s] to earth)
Mains power quality should be that
of a typical commercial or hospital
environment.
Power Frequency
30 A/m
Power frequency magnetic fields
10
Figures in parentheses are the immunity compliance levels for each test.
11
Applies only to the Merlin PCS.