User's Manual

30
Table 5. Guidance and manufacturer's declaration - electromagnetic immunity (conducted RF
and radiated RF)
Test
IEC 60601
Test Level
13
Immunity
Compliance Level
Electromagnetic Environment -
Guidance
14
80 MHz to 800 MHz
16
800 MHz to 2.7 GHz
where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey*, should
be less than the compliance level in
each frequency range.** Interference
may occur in the vicinity of equipment
marked with the following symbol:
*Electromagnetic site survey- Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should
be considered. If the measured field strength of the location in which the Merlin PSA is used
exceeds the applicable RF compliance level above, the Merlin PSA should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the Merlin PSA.
**Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than [V
1
]V/m.
NOTE: The Merlin PCS is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The user of the Merlin PCS can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
15
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
16
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.